Recall Z-1688-2021— Ortho-Clinical Diagnostics, INc.

Class II · Terminated

Class II FDA medical device recall by Ortho-Clinical Diagnostics, INc., initiated 2021-05-03, terminated 2023-10-24. FDA's recall record names no specific 510(k) clearance for this event. Product: VITROS CoV2 Ag calibrator (Product Code 6199950) - Product Usage: is a chemiluminescent immunoassay intended for the qualitative detection of SARS-CoV-2 nucleocapsid protein antigens in nasopharyngeal (NP) and anterior nasal swab specimens collected in CDCs formulation of VTM.. Reason: Ortho investigated and confirmed the potential for unexpected failed calibration, and condition codes associated with the calibration, when using VITROS SARS-CoV-2 Ag Reagent Pack along with VITROS Immunodiagnostic Products SARS-CoV-2 Ag Calibrator, Lot 20..

Recalling firm
Ortho-Clinical Diagnostics, INc.
Classification
Class II
Status
Terminated
Initiated
2021-05-03
Terminated
2023-10-24
Firm location
Rochester, NY, United States

Product description

VITROS CoV2 Ag calibrator (Product Code 6199950) - Product Usage: is a chemiluminescent immunoassay intended for the qualitative detection of SARS-CoV-2 nucleocapsid protein antigens in nasopharyngeal (NP) and anterior nasal swab specimens collected in CDCs formulation of VTM.

Reason for recall

Ortho investigated and confirmed the potential for unexpected failed calibration, and condition codes associated with the calibration, when using VITROS SARS-CoV-2 Ag Reagent Pack along with VITROS Immunodiagnostic Products SARS-CoV-2 Ag Calibrator, Lot 20.