Recall Z-1687-2017— Respironics California Inc

Class II · Terminated

Class II FDA medical device recall by Respironics California Inc, initiated 2017-02-01, terminated 2018-06-26. It names one FDA 510(k) clearance: K102985. Product: Philips V60 Ventilator with Version 2.20 Software, Description: V 60 Ventilator,Intl Opt: CFLEX,AVAPS,PPV V60 Ventilator,Intl Opt: CFLEX,AVAPS V60 Ventilator,Intl Opt: None V60 Ventilator,English Opt: None V60 Ventilator,Engl Opt: CFLEX,AVAPS V60 Us Demo Unit V60 USED ENGL OPT:CFLEX, AVAPS,PPV,AT+cl V60 VENT, JAPAN OPT: CFLEX, AVAPS, AT+ V60 VENT, JAPAN OPT: CFLEX, AVAPS, AT+ Catalog/REF No. 1053613, 1053614, 1053615, 1053616, 1053617, DU1053617, U1053617, 1076709, R1076709. Reason: The V60 Ventilator with Version 2.20 software installed may falsely detect that the blower motor has stalled. If this condition occurs, the software will cause the ventilator to shut down (Vent Inop) and display Error Code 100E. Ventilatory support will cease..

Recalling firm
Respironics California Inc
Classification
Class II
Status
Terminated
Initiated
2017-02-01
Terminated
2018-06-26
Firm location
Carlsbad, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Philips V60 Ventilator with Version 2.20 Software, Description: V 60 Ventilator,Intl Opt: CFLEX,AVAPS,PPV V60 Ventilator,Intl Opt: CFLEX,AVAPS V60 Ventilator,Intl Opt: None V60 Ventilator,English Opt: None V60 Ventilator,Engl Opt: CFLEX,AVAPS V60 Us Demo Unit V60 USED ENGL OPT:CFLEX, AVAPS,PPV,AT+cl V60 VENT, JAPAN OPT: CFLEX, AVAPS, AT+ V60 VENT, JAPAN OPT: CFLEX, AVAPS, AT+ Catalog/REF No. 1053613, 1053614, 1053615, 1053616, 1053617, DU1053617, U1053617, 1076709, R1076709

Reason for recall

The V60 Ventilator with Version 2.20 software installed may falsely detect that the blower motor has stalled. If this condition occurs, the software will cause the ventilator to shut down (Vent Inop) and display Error Code 100E. Ventilatory support will cease.