Recall Z-1685-2021— Siemens Medical Solutions USA, Inc

Class II · Terminated

Class II FDA medical device recall by Siemens Medical Solutions USA, Inc, initiated 2021-04-20, terminated 2023-11-08. It names one FDA 510(k) clearance: K152928. Product: Multitom RAX with software version VF10. Reason: Siemens identified two issues, In some cases, it is possible to perform an image acquisition although more than 20% of the beam does not hit the detector. This may only occur when an organ program for the wall mode without top alignment is selected and the system is moved to the centred position. If afterwards the operator activates the top alignment, increases the collimation, and lifts the tube, the system allows examination even with the tube being misaligned to the detector position. Hence, it is possible to overshoot the detector by more than 20%. From a clinical point of view, it is highly unlikely for the operator to trigger the examination as the misalignment is visible due to the light field being out of range. AND In very rare cases it may occur that the calculated dose value exceeds the limit of 2 Gy. This may potentially occur only when an organ program for free exposure is selected, and the system uses incorrect (too small) source-to-image distance (SID) for calculating applied dose. Hence, the calculated dose is much higher than the actual applied dose causing the buzzer, which normally notifies the operator about the exceeded limit of 2 Gy, to get activated erroneously..

Recalling firm
Siemens Medical Solutions USA, Inc
Classification
Class II
Status
Terminated
Initiated
2021-04-20
Terminated
2023-11-08
Firm location
Malvern, PA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Multitom RAX with software version VF10

Reason for recall

Siemens identified two issues, In some cases, it is possible to perform an image acquisition although more than 20% of the beam does not hit the detector. This may only occur when an organ program for the wall mode without top alignment is selected and the system is moved to the centred position. If afterwards the operator activates the top alignment, increases the collimation, and lifts the tube, the system allows examination even with the tube being misaligned to the detector position. Hence, it is possible to overshoot the detector by more than 20%. From a clinical point of view, it is highly unlikely for the operator to trigger the examination as the misalignment is visible due to the light field being out of range. AND In very rare cases it may occur that the calculated dose value exceeds the limit of 2 Gy. This may potentially occur only when an organ program for free exposure is selected, and the system uses incorrect (too small) source-to-image distance (SID) for calculating applied dose. Hence, the calculated dose is much higher than the actual applied dose causing the buzzer, which normally notifies the operator about the exceeded limit of 2 Gy, to get activated erroneously.

Data sourced from openFDA. This site is unofficial and independent of the FDA.