Recall Z-1682-2022— ICU Medical, Inc.
Class II · Ongoing
Class II FDA medical device recall by ICU Medical, Inc., initiated 2022-06-27. FDA's recall record names no specific 510(k) clearance for this event. Product: Product Description/Item: ChemoLock Vial Spike, 20mm/011-CL-80S; ChemoLock Bag Spike/CL-10; ChemoLock w/Mini Bag Spike/CL-17; 8 IN (20cm) APPX 1.0ml, EXT SET, MicroClave Clear, ChemoLock PORT, YCONN/CL3950; 31" (79 cm) Appx 3.4 ml, Admin Set w/20 Drop Integrated ChemoLock Drip Chamber, Spiros, Purple Cap, Hanger/CL3951; 7.5" (19 cm) Appx 1.5 ml, Ext Set w/ChemoLock Port, Clamp, Graduated Adapter/CL3967; 6 IN(15cm)APPX 0.82ml,EXT SET,MicroClave Clear,ChemoLock PORT,YCONN, ROTATING LUER/CL4118; ChemoLock Kit w/ 5" (13 cm) Bag Spike Adapter w/ChemoLock w/Red Cap, Vented Cap, ChemoLock Port Bag Spike/CL4158; ChemosafeLock Bag Spike/KL-BS001U3; ChemosafeLock Vial Adapter/KL-VA201U3; ChemosafeLock Vial Adapter/KL-VA202U3; ChemosafeLock Vial Adapter/KL-VA321U3; ChemosafeLock Vial Adapter/KL-VA322U3; ChemoLock Vented Vial Spike/011-CL-51; Transfer Set, PUR Yellow w/ChemoLock Port, Spiros/011-CL4155; Puerto ChemoLock/034-CL2100; ChemoLock, Punzon Univ. Vial c/ Filtro, fijacion/034-CL-70; ChemoLock, Punzon Vial Tritan c/ Filtro, 13mm/034-CL-72. Reason: Ports have a potential for unintended disconnection, inability to connect to or fully engage with injectors, due to a variation in the spring inside the Port, which may cause delay of therapy and exposure to caustic substances. Ports available as standalone connectors, or utilized as access points on vial adaptors, bag spikes, and administration sets..
- Recalling firm
- ICU Medical, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2022-06-27
- Firm location
- San Clemente, CA, United States
Product description
Product Description/Item: ChemoLock Vial Spike, 20mm/011-CL-80S; ChemoLock Bag Spike/CL-10; ChemoLock w/Mini Bag Spike/CL-17; 8 IN (20cm) APPX 1.0ml, EXT SET, MicroClave Clear, ChemoLock PORT, YCONN/CL3950; 31" (79 cm) Appx 3.4 ml, Admin Set w/20 Drop Integrated ChemoLock Drip Chamber, Spiros, Purple Cap, Hanger/CL3951; 7.5" (19 cm) Appx 1.5 ml, Ext Set w/ChemoLock Port, Clamp, Graduated Adapter/CL3967; 6 IN(15cm)APPX 0.82ml,EXT SET,MicroClave Clear,ChemoLock PORT,YCONN, ROTATING LUER/CL4118; ChemoLock Kit w/ 5" (13 cm) Bag Spike Adapter w/ChemoLock w/Red Cap, Vented Cap, ChemoLock Port Bag Spike/CL4158; ChemosafeLock Bag Spike/KL-BS001U3; ChemosafeLock Vial Adapter/KL-VA201U3; ChemosafeLock Vial Adapter/KL-VA202U3; ChemosafeLock Vial Adapter/KL-VA321U3; ChemosafeLock Vial Adapter/KL-VA322U3; ChemoLock Vented Vial Spike/011-CL-51; Transfer Set, PUR Yellow w/ChemoLock Port, Spiros/011-CL4155; Puerto ChemoLock/034-CL2100; ChemoLock, Punzon Univ. Vial c/ Filtro, fijacion/034-CL-70; ChemoLock, Punzon Vial Tritan c/ Filtro, 13mm/034-CL-72
Reason for recall
Ports have a potential for unintended disconnection, inability to connect to or fully engage with injectors, due to a variation in the spring inside the Port, which may cause delay of therapy and exposure to caustic substances. Ports available as standalone connectors, or utilized as access points on vial adaptors, bag spikes, and administration sets.
Data sourced from openFDA. This site is unofficial and independent of the FDA.