Recall Z-1680-2015— GE Healthcare
Class II · Terminated
Class II FDA medical device recall by GE Healthcare, initiated 2015-04-24, terminated 2016-09-07. It names one FDA 510(k) clearance: K051092. Product: D-fend, (p/n 876446-HEL), and D-fend+ (p/n 881319-HEL) water trap. Used with Compact Airway Modules E/M-C(Ai)O(V)(X) and Cardiocap/5 monitors. Product Usage: Water traps are intended to protect the GE monitor from humidity, secretions and bacterial contamination when monitoring respiratory parameters. D-fend and D-fend+ water traps are used with Compact gas module E-C(Ai)O(V)(X) and CardioCap/5 monitors. The Datex-Ohmeda S/5 Compact Airway Module, E-CAiOVX family is indicated for monitoring hospital patients respiration (CO2, O2, N2O, anesthetic agents, anesthetic agent identification and respiration rate) ventilation (airway pressure, volume and flow) and gas exchange status (Oxygen Consumption VO2, Carbon Dioxide production VCO2, Respiratory Quotient RQ, and Energy Expenditure, EE). Gas exchange status monitoring is not indicated in the presence of N2O+O2 mixtures. The device is indicated for use by qualified medical personnel only.. Reason: A potential safety issue due to loss of gas monitoring associated with occlusion of specific D-Fend / D-Fend+ water traps used with Compact Airway Modules E/M-C(Ai)O(V)(X) and Cardiocap/5 monitors. This water trap occlusion could cause a loss of respiratory airway gas monitoring..
- Recalling firm
- GE Healthcare
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2015-04-24
- Terminated
- 2016-09-07
- Firm location
- Waukesha, WI, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
D-fend, (p/n 876446-HEL), and D-fend+ (p/n 881319-HEL) water trap. Used with Compact Airway Modules E/M-C(Ai)O(V)(X) and Cardiocap/5 monitors. Product Usage: Water traps are intended to protect the GE monitor from humidity, secretions and bacterial contamination when monitoring respiratory parameters. D-fend and D-fend+ water traps are used with Compact gas module E-C(Ai)O(V)(X) and CardioCap/5 monitors. The Datex-Ohmeda S/5 Compact Airway Module, E-CAiOVX family is indicated for monitoring hospital patients respiration (CO2, O2, N2O, anesthetic agents, anesthetic agent identification and respiration rate) ventilation (airway pressure, volume and flow) and gas exchange status (Oxygen Consumption VO2, Carbon Dioxide production VCO2, Respiratory Quotient RQ, and Energy Expenditure, EE). Gas exchange status monitoring is not indicated in the presence of N2O+O2 mixtures. The device is indicated for use by qualified medical personnel only.
Reason for recall
A potential safety issue due to loss of gas monitoring associated with occlusion of specific D-Fend / D-Fend+ water traps used with Compact Airway Modules E/M-C(Ai)O(V)(X) and Cardiocap/5 monitors. This water trap occlusion could cause a loss of respiratory airway gas monitoring.