Recall Z-1678-2025— Belmont Instrument LLC

Class II · Ongoing

Class II FDA medical device recall by Belmont Instrument LLC, initiated 2025-03-31. It names one FDA 510(k) clearance: K242735. Product: 3.0 Liter Disposable Set (4 sets/case). Model Number: 903-00032. The P/N 903-00032 consists of 4 Large Volume Reservoirs (P/N 903-00018) and 4 Heat Exchanger/Patient Line set (P/N 403-00382). The two components are packaged separately in individual pouches and placed in the same case for shipment. The female quick connector that may be cracked is located on the Heat Exchanger/Patient Line set.. Reason: Potential crack in the female connector located on the Heat Exchanger/patient line set, which will lead to a fluid leak through the crack during priming of the Rapid Infuser..

Recalling firm
Belmont Instrument LLC
Classification
Class II
Status
Ongoing
Initiated
2025-03-31
Firm location
Billerica, MA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

3.0 Liter Disposable Set (4 sets/case). Model Number: 903-00032. The P/N 903-00032 consists of 4 Large Volume Reservoirs (P/N 903-00018) and 4 Heat Exchanger/Patient Line set (P/N 403-00382). The two components are packaged separately in individual pouches and placed in the same case for shipment. The female quick connector that may be cracked is located on the Heat Exchanger/Patient Line set.

Reason for recall

Potential crack in the female connector located on the Heat Exchanger/patient line set, which will lead to a fluid leak through the crack during priming of the Rapid Infuser.