Recall Z-1678-2020— Siemens Medical Solutions USA, Inc

Class II · Terminated

Class II FDA medical device recall by Siemens Medical Solutions USA, Inc, initiated 2020-02-18, terminated 2021-02-09. It names 2 FDA 510(k) clearances: K982513, K882729. Product: MEVATRON M2/Primus Mid-Energy Model #01940035, MEVATRON PRIMART #05500371, MEVATRON KD-2 #08515520, MEVATRON KD2 #01940753, MEVATRON M-2 6740 #09401506, MEVATRON MD-2 #09401654, MEVATRON MDX-2 #09401746 with Digital LINAC Systems Control Console Software Versions 6, 7, 9, 13. Reason: Potential safety issue with the dose monitoring system safety interlocks which could result in a 10-20% overdose of radiation therapy..

Recalling firm
Siemens Medical Solutions USA, Inc
Classification
Class II
Status
Terminated
Initiated
2020-02-18
Terminated
2021-02-09
Firm location
Malvern, PA, United States

510(k) clearances named in this recall

2 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

MEVATRON M2/Primus Mid-Energy Model #01940035, MEVATRON PRIMART #05500371, MEVATRON KD-2 #08515520, MEVATRON KD2 #01940753, MEVATRON M-2 6740 #09401506, MEVATRON MD-2 #09401654, MEVATRON MDX-2 #09401746 with Digital LINAC Systems Control Console Software Versions 6, 7, 9, 13

Reason for recall

Potential safety issue with the dose monitoring system safety interlocks which could result in a 10-20% overdose of radiation therapy.