Recall Z-1673-2023— Smith & Nephew, Inc.

Class II · Ongoing

Class II FDA medical device recall by Smith & Nephew, Inc., initiated 2023-05-02. It names one FDA 510(k) clearance: K190439. Product: Smith&Nephew ENGAGE Cementless Partial Knee System, Coated Tibial Insert, Part Numbers: a) REF 1-10013-109, SIZE 1-LEFT MED, 9MM; b) REF 1-10013-110, SIZE 1-LEFT MED, 10MM; c) REF 1-10013-111, SIZE 1-LEFT MED, 11MM; d) REF 1-10013-113, SIZE 1-LEFT MED, 13MM; e) REF 1-10013-159, SIZE 1-RIGHT MED, 9MM; f) REF 1-10013-160, SIZE 1-RIGHT MED, 10MM; g) REF 1-10013-161, SIZE 1-RIGHT MED, 11MM; h) REF 1-10013-163, SIZE 1-RIGHT MED, 13MM; i) REF 1-10013-209, SIZE 2-LEFT MED, 9MM; j) REF 1-10013-210, SIZE 2-LEFT MED, 10MM; k) REF 1-10013-211, SIZE 2-LEFT MED, 11MM; l) REF 1-10013-213, SIZE 2-LEFT MED, 13MM; m) REF 1-10013-259, SIZE 2-RIGHT MED, 9MM; n) REF 1-10013-260, SIZE 2-RIGHT MED, 10MM; o) REF 1-10013-261, SIZE 2-RIGHT MED, 11MM; p) REF 1-10013-263, SIZE 2-RIGHT MED, 13MM; q) REF 1-10013-309, SIZE 3-LEFT MED, 9MM; r) REF 1-10013-310, SIZE 3-LEFT MED, 10MM; s) REF 1-10013-311, SIZE 3-LEFT MED, 11MM; t) REF 1-10013-313, SIZE 3-LEFT MED, 13MM; u) REF 1-10013-359, SIZE 3-RIGHT MED, 9MM; v) REF 1-10013-360, SIZE 3-RIGHT MED, 10MM; w) REF 1-10013-361, SIZE 3-RIGHT MED, 11MM; x) REF 1-10013-363, SIZE 3-RIGHT MED, 13MM; y) REF 1-10013-409, SIZE 4-LEFT MED, 9MM; z) REF 1-10013-410, SIZE 4-LEFT MED, 10MM; aa) REF 1-10013-411, SIZE 4-LEFT MED, 11MM; bb) REF 1-10013-413, SIZE 4-LEFT MED, 13MM; cc) REF 1-10013-459, SIZE 4-RIGHT MED, 9MM; dd) REF 1-10013-460, SIZE 4-RIGHT MED, 10MM; ee) REF 1-10013-461, SIZE 4-RIGHT MED, 11MM; ff) REF 1-10013-463, SIZE 4-RIGHT MED, 13MM; gg) REF 1-10013-509, SIZE 5-LEFT MED, 9MM; hh) REF 1-10013-510, SIZE 5-LEFT MED, 10MM; ii) REF 1-10013-511, SIZE 5-LEFT MED, 11MM; jj) REF 1-10013-513, SIZE 5-LEFT MED, 13MM; kk) REF 1-10013-559, SIZE 5-RIGHT MED, 9MM; ll) REF 1-10013-560, SIZE 5-RIGHT MED, 10MM; mm) REF 1-10013-561, SIZE 5-RIGHT MED, 11MM; nn) REF 1-10013-563, SIZE 5-RIGHT MED, 13MM; oo) REF 1-10013-609, SIZE 6-LEFT MED, 9MM; pp) REF 1-10013-610, SIZE 6-LEFT MED, 10MM; qq) REF 1-10013-611, SIZE 6-LEFT MED, 11MM; rr) REF 1-10013-613, SIZE 6-LEFT MED, 13MM; ss) REF 1-10013-659, SIZE 6-RIGHT MED, 9MM; tt) REF 1-10013-660, SIZE 6-RIGHT MED, 10MM; uu) REF 1-10013-661, SIZE 6-RIGHT MED, 11MM; vv) REF 1-10013-663, SIZE 6-RIGHT MED, 13MM; Unicondylar knee prothesis. Reason: Recent complaint data indicates that the revision rate may be trending higher than corresponding similar devices in global joint replacement registries. The data identifies a potential signal that the performance is an outlier versus the state of the art with respect to the risk for revision..

Recalling firm
Smith & Nephew, Inc.
Classification
Class II
Status
Ongoing
Initiated
2023-05-02
Firm location
Memphis, TN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Smith&Nephew ENGAGE Cementless Partial Knee System, Coated Tibial Insert, Part Numbers: a) REF 1-10013-109, SIZE 1-LEFT MED, 9MM; b) REF 1-10013-110, SIZE 1-LEFT MED, 10MM; c) REF 1-10013-111, SIZE 1-LEFT MED, 11MM; d) REF 1-10013-113, SIZE 1-LEFT MED, 13MM; e) REF 1-10013-159, SIZE 1-RIGHT MED, 9MM; f) REF 1-10013-160, SIZE 1-RIGHT MED, 10MM; g) REF 1-10013-161, SIZE 1-RIGHT MED, 11MM; h) REF 1-10013-163, SIZE 1-RIGHT MED, 13MM; i) REF 1-10013-209, SIZE 2-LEFT MED, 9MM; j) REF 1-10013-210, SIZE 2-LEFT MED, 10MM; k) REF 1-10013-211, SIZE 2-LEFT MED, 11MM; l) REF 1-10013-213, SIZE 2-LEFT MED, 13MM; m) REF 1-10013-259, SIZE 2-RIGHT MED, 9MM; n) REF 1-10013-260, SIZE 2-RIGHT MED, 10MM; o) REF 1-10013-261, SIZE 2-RIGHT MED, 11MM; p) REF 1-10013-263, SIZE 2-RIGHT MED, 13MM; q) REF 1-10013-309, SIZE 3-LEFT MED, 9MM; r) REF 1-10013-310, SIZE 3-LEFT MED, 10MM; s) REF 1-10013-311, SIZE 3-LEFT MED, 11MM; t) REF 1-10013-313, SIZE 3-LEFT MED, 13MM; u) REF 1-10013-359, SIZE 3-RIGHT MED, 9MM; v) REF 1-10013-360, SIZE 3-RIGHT MED, 10MM; w) REF 1-10013-361, SIZE 3-RIGHT MED, 11MM; x) REF 1-10013-363, SIZE 3-RIGHT MED, 13MM; y) REF 1-10013-409, SIZE 4-LEFT MED, 9MM; z) REF 1-10013-410, SIZE 4-LEFT MED, 10MM; aa) REF 1-10013-411, SIZE 4-LEFT MED, 11MM; bb) REF 1-10013-413, SIZE 4-LEFT MED, 13MM; cc) REF 1-10013-459, SIZE 4-RIGHT MED, 9MM; dd) REF 1-10013-460, SIZE 4-RIGHT MED, 10MM; ee) REF 1-10013-461, SIZE 4-RIGHT MED, 11MM; ff) REF 1-10013-463, SIZE 4-RIGHT MED, 13MM; gg) REF 1-10013-509, SIZE 5-LEFT MED, 9MM; hh) REF 1-10013-510, SIZE 5-LEFT MED, 10MM; ii) REF 1-10013-511, SIZE 5-LEFT MED, 11MM; jj) REF 1-10013-513, SIZE 5-LEFT MED, 13MM; kk) REF 1-10013-559, SIZE 5-RIGHT MED, 9MM; ll) REF 1-10013-560, SIZE 5-RIGHT MED, 10MM; mm) REF 1-10013-561, SIZE 5-RIGHT MED, 11MM; nn) REF 1-10013-563, SIZE 5-RIGHT MED, 13MM; oo) REF 1-10013-609, SIZE 6-LEFT MED, 9MM; pp) REF 1-10013-610, SIZE 6-LEFT MED, 10MM; qq) REF 1-10013-611, SIZE 6-LEFT MED, 11MM; rr) REF 1-10013-613, SIZE 6-LEFT MED, 13MM; ss) REF 1-10013-659, SIZE 6-RIGHT MED, 9MM; tt) REF 1-10013-660, SIZE 6-RIGHT MED, 10MM; uu) REF 1-10013-661, SIZE 6-RIGHT MED, 11MM; vv) REF 1-10013-663, SIZE 6-RIGHT MED, 13MM; Unicondylar knee prothesis

Reason for recall

Recent complaint data indicates that the revision rate may be trending higher than corresponding similar devices in global joint replacement registries. The data identifies a potential signal that the performance is an outlier versus the state of the art with respect to the risk for revision.

Data sourced from openFDA. This site is unofficial and independent of the FDA.