Recall Z-1670-2015— Siemens Medical Solutions USA, Inc
Class II · Terminated
Class II FDA medical device recall by Siemens Medical Solutions USA, Inc, initiated 2015-04-13, terminated 2017-03-20. It names one FDA 510(k) clearance: K101491. Product: SIEMENS Uroskop Omnia Max ; a solid state detector fluoroscopic X-Ray system, primarily for urological applications (functional x-ray diagnostic, endourology and minimal invasive urology/surgery). The system,, which includes a radiologic/urologic treatment table, may be used for urological, gastroenterological and gynecological treatment, planning and diagnostic procedures including but not limited to: Querying and retrieving patient history information and/or previous diagnosis and images from other modalities, including X-ray examinations of the urogenital area, Ultrasound examinations, Endourological interventions, Percutaneous interventions, Laparoscopy, Application of fistula, Simple procedures, Extracorporeal shock wave lithotripsy, Uroflow/urodynamics, Pediatric radiological and therapeutic applications.. Reason: Study and all acquired images deleted when using systems with software version VE10E. Acquisitions have to be repeated. Connection with portable detectors (MAX wi-D, MAX mini) is sporadic in image recovery. Pressing undo button on the RAD subtask card or previous series button on the Image subtask card during image readout. Sporadically, during an automatic or a manual RIS update..
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2015-04-13
- Terminated
- 2017-03-20
- Firm location
- Malvern, PA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
SIEMENS Uroskop Omnia Max ; a solid state detector fluoroscopic X-Ray system, primarily for urological applications (functional x-ray diagnostic, endourology and minimal invasive urology/surgery). The system,, which includes a radiologic/urologic treatment table, may be used for urological, gastroenterological and gynecological treatment, planning and diagnostic procedures including but not limited to: Querying and retrieving patient history information and/or previous diagnosis and images from other modalities, including X-ray examinations of the urogenital area, Ultrasound examinations, Endourological interventions, Percutaneous interventions, Laparoscopy, Application of fistula, Simple procedures, Extracorporeal shock wave lithotripsy, Uroflow/urodynamics, Pediatric radiological and therapeutic applications.
Reason for recall
Study and all acquired images deleted when using systems with software version VE10E. Acquisitions have to be repeated. Connection with portable detectors (MAX wi-D, MAX mini) is sporadic in image recovery. Pressing undo button on the RAD subtask card or previous series button on the Image subtask card during image readout. Sporadically, during an automatic or a manual RIS update.