Recall Z-1664-2025— Beckman Coulter, Inc.

Class II · Ongoing

Class II FDA medical device recall by Beckman Coulter, Inc., initiated 2025-03-27. It names one FDA 510(k) clearance: K221225. Product: Beckman Coulter DxI 9000 Access Immunoassay Analyzer. Reason: When DxI 9000 Access Immunoassay Analyzer is connected to a host system (laboratory information system (LIS) or middleware) and has accumulated canceled QC test results that were not sent by LIS, the analyzer may lose communication with the host system. The lost connection interrupts sample processing, which delays reporting patient test results and may subsequently delay patient treatment..

Recalling firm
Beckman Coulter, Inc.
Classification
Class II
Status
Ongoing
Initiated
2025-03-27
Firm location
Chaska, MN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Beckman Coulter DxI 9000 Access Immunoassay Analyzer

Reason for recall

When DxI 9000 Access Immunoassay Analyzer is connected to a host system (laboratory information system (LIS) or middleware) and has accumulated canceled QC test results that were not sent by LIS, the analyzer may lose communication with the host system. The lost connection interrupts sample processing, which delays reporting patient test results and may subsequently delay patient treatment.