Recall Z-1663-2025— IHB OPERATIONS B.V.

Class II · Ongoing

Class II FDA medical device recall by IHB OPERATIONS B.V., initiated 2025-03-27. FDA's recall record names no specific 510(k) clearance for this event. Product: Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount. Reason: The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column..

Recalling firm
IHB OPERATIONS B.V.
Classification
Class II
Status
Ongoing
Initiated
2025-03-27
Firm location
Zwolle, N/A, Netherlands

Product description

Skytron Freedom 4FXS-60 Heavy Duty Monitor Mount

Reason for recall

The mounting hardware that is supplied with the 4FXS-60 monitor bracket does not provide adequate thread engagement to support the weight of the frame and monitor. This could lead to a failure of the mounting hardware and subsequent separation of the frame from the mounting column.