Recall Z-1660-2014— Physio-Control, Inc.

Class II · Terminated

Class II FDA medical device recall by Physio-Control, Inc., initiated 2014-05-09, terminated 2018-05-29. It names one FDA 510(k) clearance: K042404. Product: LIFEPAK 1000 defibrillator. Affected Part Numbers: 320371500XXX The LP1000 (in AED mode) is indicated for use on patients in cardiac arrest.. Reason: Physio-Control has become aware of incidents where customers have attempted to use their LIFEPAK 1000 defibrillator and the device has shut down unexpectedly due to a very low battery. A defibrillator in this scenario has the potential to fail to deliver a shock, with the potential result that therapy is not delivered and a patient is not resuscitated. A software malfunction in the LIFEPAK 100.

Recalling firm
Physio-Control, Inc.
Classification
Class II
Status
Terminated
Initiated
2014-05-09
Terminated
2018-05-29
Firm location
Redmond, WA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

LIFEPAK 1000 defibrillator. Affected Part Numbers: 320371500XXX The LP1000 (in AED mode) is indicated for use on patients in cardiac arrest.

Reason for recall

Physio-Control has become aware of incidents where customers have attempted to use their LIFEPAK 1000 defibrillator and the device has shut down unexpectedly due to a very low battery. A defibrillator in this scenario has the potential to fail to deliver a shock, with the potential result that therapy is not delivered and a patient is not resuscitated. A software malfunction in the LIFEPAK 100