Recall Z-1659-2018— Siemens Healthcare Diagnostics, Inc.

Class II · Terminated

Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc., initiated 2017-06-29, terminated 2020-04-23. It names one FDA 510(k) clearance: K083465. Product: Dimension Vista Creatine Kinase MB (MBI) assay; (Product Number: K3032/10464339) (Mfr: Siemens Healthcare Diagnostics, Newark, DE). Reason: Various assays used to measure different chemical levels in blood, urine, or cerebrospinal fluid may interfere with the drugs sulfasalazine and/or sulfapyridine causing falsely elevated or falsely depressed results..

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Classification
Class II
Status
Terminated
Initiated
2017-06-29
Terminated
2020-04-23
Firm location
Tarrytown, NY, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Dimension Vista Creatine Kinase MB (MBI) assay; (Product Number: K3032/10464339) (Mfr: Siemens Healthcare Diagnostics, Newark, DE)

Reason for recall

Various assays used to measure different chemical levels in blood, urine, or cerebrospinal fluid may interfere with the drugs sulfasalazine and/or sulfapyridine causing falsely elevated or falsely depressed results.

Data sourced from openFDA. This site is unofficial and independent of the FDA.