Recall Z-1658-2023— Maquet Medical Systems USA

Class II · Ongoing

Class II FDA medical device recall by Maquet Medical Systems USA, initiated 2023-05-18. It names one FDA 510(k) clearance: K100278. Product: QUADROX-iD Pediatric, Oxygenator used during cardiopulmonary bypass Model Item No. BEQ-HMOD30000-USA 701070397 BEQ-HMOD30000-USA 701050330. Reason: Product packaging may exhibit small pinholes which may not be visible to the unaided eye, compromising the sterile barrier of the packaging. Exposure to non-sterile device may result in inflammation, infection, sepsis, and ischemia..

Recalling firm
Maquet Medical Systems USA
Classification
Class II
Status
Ongoing
Initiated
2023-05-18
Firm location
Wayne, NJ, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

QUADROX-iD Pediatric, Oxygenator used during cardiopulmonary bypass Model Item No. BEQ-HMOD30000-USA 701070397 BEQ-HMOD30000-USA 701050330

Reason for recall

Product packaging may exhibit small pinholes which may not be visible to the unaided eye, compromising the sterile barrier of the packaging. Exposure to non-sterile device may result in inflammation, infection, sepsis, and ischemia.

Data sourced from openFDA. This site is unofficial and independent of the FDA.