Recall Z-1656-2026— Reflexion Medical, Inc.

Class II · Ongoing

Class II FDA medical device recall by Reflexion Medical, Inc., initiated 2026-02-09. FDA's recall record names no specific 510(k) clearance for this event. Product: RefleXion X1 Radiotherapy System Model RXM1000, Part #800-00006-00. Reason: Due to a radiotherapy medical system software defect, when performing angular roll corrections followed by a repeat localization, the roll corrections applied in the initial localization scan are not carried through to treatment delivery leading to the treatment plan dose being delivered to the incorrect location, potentially a displacement of 5mm or higher, with a 10% to 20% underdose..

Recalling firm
Reflexion Medical, Inc.
Classification
Class II
Status
Ongoing
Initiated
2026-02-09
Firm location
Hayward, CA, United States

Product description

RefleXion X1 Radiotherapy System Model RXM1000, Part #800-00006-00

Reason for recall

Due to a radiotherapy medical system software defect, when performing angular roll corrections followed by a repeat localization, the roll corrections applied in the initial localization scan are not carried through to treatment delivery leading to the treatment plan dose being delivered to the incorrect location, potentially a displacement of 5mm or higher, with a 10% to 20% underdose.