Recall Z-1656-2025— Abbott Laboratories

Class II · Ongoing

Class II FDA medical device recall by Abbott Laboratories, initiated 2025-03-18. It names one FDA 510(k) clearance: K061667. Product: Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 04U42-84 Software Version: All software versions Product Description: CELL-DYN Ruby. Reason: When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual..

Recalling firm
Abbott Laboratories
Classification
Class II
Status
Ongoing
Initiated
2025-03-18
Firm location
Abbott Park, IL, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 04U42-84 Software Version: All software versions Product Description: CELL-DYN Ruby

Reason for recall

When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual.

Data sourced from openFDA. This site is unofficial and independent of the FDA.