Recall Z-1654-2022— Philips Respironics, Inc.
Class I · Ongoing
Class I FDA medical device recall by Philips Respironics, Inc., initiated 2022-08-26. It names one FDA 510(k) clearance: K113053. Product: Philips BiPAP A30, Rx Only. To provide noninvasive ventilation support for patients with obstructive sleep apnea syndrome (OSA) and respiratory insufficiency.. Reason: Motor assemblies manufactured with nonconforming plastic that contributes to off-gassing of VOCs..
- Recalling firm
- Philips Respironics, Inc.
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2022-08-26
- Firm location
- Murrysville, PA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Philips BiPAP A30, Rx Only. To provide noninvasive ventilation support for patients with obstructive sleep apnea syndrome (OSA) and respiratory insufficiency.
Reason for recall
Motor assemblies manufactured with nonconforming plastic that contributes to off-gassing of VOCs.