Recall Z-1652-2018— Siemens Healthcare Diagnostics, Inc.

Class II · Terminated

Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc., initiated 2017-06-29, terminated 2020-04-23. FDA's recall record names no specific 510(k) clearance for this event. Product: Dimension Glucose (GLUC) assay (Product Number: DF40/10444973) (Mfr: Siemens Healthcare Diagnostics, Newark, DE). Reason: Various assays used to measure different chemical levels in blood, urine, or cerebrospinal fluid may interfere with the drugs sulfasalazine and/or sulfapyridine causing falsely elevated or falsely depressed results..

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Classification
Class II
Status
Terminated
Initiated
2017-06-29
Terminated
2020-04-23
Firm location
Tarrytown, NY, United States

Product description

Dimension Glucose (GLUC) assay (Product Number: DF40/10444973) (Mfr: Siemens Healthcare Diagnostics, Newark, DE)

Reason for recall

Various assays used to measure different chemical levels in blood, urine, or cerebrospinal fluid may interfere with the drugs sulfasalazine and/or sulfapyridine causing falsely elevated or falsely depressed results.

Data sourced from openFDA. This site is unofficial and independent of the FDA.