Recall Z-1651-2023— Philips Respironics, Inc.

Class I · Ongoing

Class I FDA medical device recall by Philips Respironics, Inc., initiated 2023-05-01. It names one FDA 510(k) clearance: K181166. Product: Trilogy Evo O2 EV300, Model Numbers DS2200X11B, IN2200X15B, FX2200X15B, CA2200X12B, FR2200X14B, ND2200X15B, ES2200X15B, BL2200X15B, GB2200X15B, EU2200X15B, EE2200X15B, and TR2200X15B. Reason: Environmental debris may accumulate on the internal machine flow sensor causing partial occlusion which may impact accurate delivery of pressure, volume, or flow. A malfunction caused by this issue can result in patient harm up to hypoxemia if not addressed by the care provider..

Recalling firm
Philips Respironics, Inc.
Classification
Class I
Status
Ongoing
Initiated
2023-05-01
Firm location
Murrysville, PA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Trilogy Evo O2 EV300, Model Numbers DS2200X11B, IN2200X15B, FX2200X15B, CA2200X12B, FR2200X14B, ND2200X15B, ES2200X15B, BL2200X15B, GB2200X15B, EU2200X15B, EE2200X15B, and TR2200X15B

Reason for recall

Environmental debris may accumulate on the internal machine flow sensor causing partial occlusion which may impact accurate delivery of pressure, volume, or flow. A malfunction caused by this issue can result in patient harm up to hypoxemia if not addressed by the care provider.

Data sourced from openFDA. This site is unofficial and independent of the FDA.