Recall Z-1649-2023— Philips Respironics, Inc.
Class I · Ongoing
Class I FDA medical device recall by Philips Respironics, Inc., initiated 2023-05-01. It names one FDA 510(k) clearance: K181166. Product: Trilogy Evo, Garbin Evo, and Aeris EVO, Model Numbers DS2110X11B, IN2110X15B, JP2110X16B, AU2110X15B, LA2110X15B, CA2110X12B, BR2110X18B, KR2110X15B, IA2110X15B, FP2110X10, FR2110X14B, ND2110X15B, IT2110X21B, ES2110X15B, DE2110X13B, BL2110X15B, GB2110X15B, EU2110X15B, EE2110X15B, LD2110X23B, and VT2110X24B. Reason: Environmental debris may accumulate on the internal machine flow sensor causing partial occlusion which may impact accurate delivery of pressure, volume, or flow. A malfunction caused by this issue can result in patient harm up to hypoxemia if not addressed by the care provider..
- Recalling firm
- Philips Respironics, Inc.
- Classification
- Class I
- Status
- Ongoing
- Initiated
- 2023-05-01
- Firm location
- Murrysville, PA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Trilogy Evo, Garbin Evo, and Aeris EVO, Model Numbers DS2110X11B, IN2110X15B, JP2110X16B, AU2110X15B, LA2110X15B, CA2110X12B, BR2110X18B, KR2110X15B, IA2110X15B, FP2110X10, FR2110X14B, ND2110X15B, IT2110X21B, ES2110X15B, DE2110X13B, BL2110X15B, GB2110X15B, EU2110X15B, EE2110X15B, LD2110X23B, and VT2110X24B
Reason for recall
Environmental debris may accumulate on the internal machine flow sensor causing partial occlusion which may impact accurate delivery of pressure, volume, or flow. A malfunction caused by this issue can result in patient harm up to hypoxemia if not addressed by the care provider.
Data sourced from openFDA. This site is unofficial and independent of the FDA.