Recall Z-1649-2018— Siemens Healthcare Diagnostics, Inc.
Class II · Terminated
Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc., initiated 2017-06-29, terminated 2020-04-23. It names one FDA 510(k) clearance: K123320. Product: Dimension Ammonia (AMM) assay; (Product Number: DF119/10711991) (Mfr: Siemens Healthcare Diagnostics, Newark, DE). Reason: Various assays used to measure different chemical levels in blood, urine, or cerebrospinal fluid may interfere with the drugs sulfasalazine and/or sulfapyridine causing falsely elevated or falsely depressed results..
- Recalling firm
- Siemens Healthcare Diagnostics, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2017-06-29
- Terminated
- 2020-04-23
- Firm location
- Tarrytown, NY, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Dimension Ammonia (AMM) assay; (Product Number: DF119/10711991) (Mfr: Siemens Healthcare Diagnostics, Newark, DE)
Reason for recall
Various assays used to measure different chemical levels in blood, urine, or cerebrospinal fluid may interfere with the drugs sulfasalazine and/or sulfapyridine causing falsely elevated or falsely depressed results.
Data sourced from openFDA. This site is unofficial and independent of the FDA.