Recall Z-1646-2018— Siemens Healthcare Diagnostics, Inc.

Class II · Terminated

Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc., initiated 2017-06-29, terminated 2020-04-23. It names one FDA 510(k) clearance: K023841. Product: ADVIA Ammonia (AMM) chemistry assay, an in vitro diagnostic device intended to quantitatively measure Ammonia levels in human plasma (heparin or EDTA). Product Number: 10286035 (Mfr: Randox Laboratories). Reason: Various assays used to measure different chemical levels in blood, urine, or cerebrospinal fluid may interfere with the drugs sulfasalazine and/or sulfapyridine causing falsely elevated or falsely depressed results..

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Classification
Class II
Status
Terminated
Initiated
2017-06-29
Terminated
2020-04-23
Firm location
Tarrytown, NY, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

ADVIA Ammonia (AMM) chemistry assay, an in vitro diagnostic device intended to quantitatively measure Ammonia levels in human plasma (heparin or EDTA). Product Number: 10286035 (Mfr: Randox Laboratories)

Reason for recall

Various assays used to measure different chemical levels in blood, urine, or cerebrospinal fluid may interfere with the drugs sulfasalazine and/or sulfapyridine causing falsely elevated or falsely depressed results.