Recall Z-1645-2025— Zimmer, Inc.

Class II · Ongoing

Class II FDA medical device recall by Zimmer, Inc., initiated 2025-03-24. It names one FDA 510(k) clearance: K181171. Product: Zimmer Biomet Ceramic Head, 22.2 mm Diameter, 12/14 Taper, +0 Neck, Model Number 802602202; Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non- Porous, Uncemented. Reason: The associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems..

Recalling firm
Zimmer, Inc.
Classification
Class II
Status
Ongoing
Initiated
2025-03-24
Firm location
Warsaw, IN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Zimmer Biomet Ceramic Head, 22.2 mm Diameter, 12/14 Taper, +0 Neck, Model Number 802602202; Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented or Non- Porous, Uncemented

Reason for recall

The associated product labeling does not adequately provide compatibility information for use of the 22.2 mm Diameter Ceramic Femoral Heads with cobalt-chrome femoral stems or stainless-steel femoral stems.

Data sourced from openFDA. This site is unofficial and independent of the FDA.