Recall Z-1641-2018— Medtronic Sofamor Danek USA Inc

Class II · Terminated

Class II FDA medical device recall by Medtronic Sofamor Danek USA Inc, initiated 2018-03-13, terminated 2020-11-02. FDA's recall record names no specific 510(k) clearance for this event. Product: Medtronic Prestige LP(TM) Cervical Disc System with Streamlined Instruments, DRILL GUIDES: a) REF 6975205, 5mm b) REF 6975206, 6mm c) REF 6975207, 7mm Product Usage: The PRESTIGE LP reusable instruments are intended for the surgical implantation of the PRESTIGE LP Cervical Disc. Medtronic reusable instruments must be sterilized prior to initial use and must be cleaned and sterilized prior to each re-use.. Reason: The firm received complaints of drill bits breaking during use with the Prestige LP(TM) Cervical Disc System. Subsequent investigation demonstrated the interaction of the drill guide and the drill bit may lead to drill bit breakage..

Recalling firm
Medtronic Sofamor Danek USA Inc
Classification
Class II
Status
Terminated
Initiated
2018-03-13
Terminated
2020-11-02
Firm location
Memphis, TN, United States

Product description

Medtronic Prestige LP(TM) Cervical Disc System with Streamlined Instruments, DRILL GUIDES: a) REF 6975205, 5mm b) REF 6975206, 6mm c) REF 6975207, 7mm Product Usage: The PRESTIGE LP reusable instruments are intended for the surgical implantation of the PRESTIGE LP Cervical Disc. Medtronic reusable instruments must be sterilized prior to initial use and must be cleaned and sterilized prior to each re-use.

Reason for recall

The firm received complaints of drill bits breaking during use with the Prestige LP(TM) Cervical Disc System. Subsequent investigation demonstrated the interaction of the drill guide and the drill bit may lead to drill bit breakage.

Data sourced from openFDA. This site is unofficial and independent of the FDA.