Recall Z-1640-2014— GE Healthcare, LLC
Class II · Terminated
Class II FDA medical device recall by GE Healthcare, LLC, initiated 2014-04-09, terminated 2015-09-28. It names 2 FDA 510(k) clearances: K051635, K955026. Product: GE Healthcare, NeuroMuscular Transmission Module, E-NMT-01 Product Usage: The Datex-Ohmeda S/5TM E-NMT module is indicated for monitoring the relaxation of the patient and regional block stimulation for nerve location.. Reason: When the E-NMT-01 module is used in conjunction with the ElectroSensor, the Neuromuscular Transmission (NMT) values may indicate a deeper level of muscle relaxation than the actual level of muscle relaxation. In the clinical situation visual movements of the hand are seen after TOF (Train of Four) stimulation, but the patient monitor shows no counts, or counts are not corresponding to the actual a.
- Recalling firm
- GE Healthcare, LLC
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2014-04-09
- Terminated
- 2015-09-28
- Firm location
- Waukesha, WI, United States
510(k) clearances named in this recall
2 on FDA's record- K051635 — DATEX-OHMEDA S/5 NEUROMUSCULAR TRANSMISSION MODULE, E-NMT AND ACCESSORIES
- K955026 — AS/3 NMT MODULE
K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
GE Healthcare, NeuroMuscular Transmission Module, E-NMT-01 Product Usage: The Datex-Ohmeda S/5TM E-NMT module is indicated for monitoring the relaxation of the patient and regional block stimulation for nerve location.
Reason for recall
When the E-NMT-01 module is used in conjunction with the ElectroSensor, the Neuromuscular Transmission (NMT) values may indicate a deeper level of muscle relaxation than the actual level of muscle relaxation. In the clinical situation visual movements of the hand are seen after TOF (Train of Four) stimulation, but the patient monitor shows no counts, or counts are not corresponding to the actual a