Recall Z-1639-2023— Maquet Medical Systems USA

Class II · Ongoing

Class II FDA medical device recall by Maquet Medical Systems USA, initiated 2023-05-18. It names one FDA 510(k) clearance: K090511. Product: BEQ-HMO 71000-USA /QUADROX-i Adult, Oxygenator used during cardiopulmonary bypass Item no. 701067829. Reason: 1. An accessory may be improperly placed during packaging, causing creases and damage to the accessory's sterile pouch. 2. Product packaging may exhibit holes, cracks, dents, and crushed areas. Both issues may compromise the sterile barrier of the packaging, and exposure to non-sterile product may result in inflammation, infection, sepsis, and ischemia..

Recalling firm
Maquet Medical Systems USA
Classification
Class II
Status
Ongoing
Initiated
2023-05-18
Firm location
Wayne, NJ, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

BEQ-HMO 71000-USA /QUADROX-i Adult, Oxygenator used during cardiopulmonary bypass Item no. 701067829

Reason for recall

1. An accessory may be improperly placed during packaging, causing creases and damage to the accessory's sterile pouch. 2. Product packaging may exhibit holes, cracks, dents, and crushed areas. Both issues may compromise the sterile barrier of the packaging, and exposure to non-sterile product may result in inflammation, infection, sepsis, and ischemia.

Data sourced from openFDA. This site is unofficial and independent of the FDA.