Recall Z-1637-2020— Bard Peripheral Vascular Inc

Class II · Terminated

Class II FDA medical device recall by Bard Peripheral Vascular Inc, initiated 2019-04-12, terminated 2021-07-28. It names one FDA 510(k) clearance: K182796. Product: BD WAVELINQ 4F EndoAVF System, Arterial Catheter 50 cm, Venous Catheter 43 cm, Sterile EO, Rx only, Mfr. Bard Peripheral Vascular, Inc., Ref W04200, Lot S0053, UDI (01)00801741182754(17)210215(10)S0053 - Product Usage: The WavelinQ 4F EndoAVF System in indicated for the creation of an arteriovenous fistula (AVF) using concomitant ulnar artery and ulnar vein or concomitant radial artery and radial vein in patients with minimum artery and vein diameters of 2.0 mm at the fistula creation site who have chronic kidney disease and need hemodialysis.. Reason: The firm has become aware of a potential problem with their venous and arterial magnetic catheters may be at risk of experiencing a magnetic deficiency, which may result in the magnets failing to attract to one another. This may result in procedural delay in obtaining or creating a functional fistula and need for additional contrast to perform the necessary fluoroscopy. This can pose an incremental risk of harm to a diabetic patient's kidney function..

Recalling firm
Bard Peripheral Vascular Inc
Classification
Class II
Status
Terminated
Initiated
2019-04-12
Terminated
2021-07-28
Firm location
Tempe, AZ, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

BD WAVELINQ 4F EndoAVF System, Arterial Catheter 50 cm, Venous Catheter 43 cm, Sterile EO, Rx only, Mfr. Bard Peripheral Vascular, Inc., Ref W04200, Lot S0053, UDI (01)00801741182754(17)210215(10)S0053 - Product Usage: The WavelinQ 4F EndoAVF System in indicated for the creation of an arteriovenous fistula (AVF) using concomitant ulnar artery and ulnar vein or concomitant radial artery and radial vein in patients with minimum artery and vein diameters of 2.0 mm at the fistula creation site who have chronic kidney disease and need hemodialysis.

Reason for recall

The firm has become aware of a potential problem with their venous and arterial magnetic catheters may be at risk of experiencing a magnetic deficiency, which may result in the magnets failing to attract to one another. This may result in procedural delay in obtaining or creating a functional fistula and need for additional contrast to perform the necessary fluoroscopy. This can pose an incremental risk of harm to a diabetic patient's kidney function.

Data sourced from openFDA. This site is unofficial and independent of the FDA.