Recall Z-1637-2016— Bradshaw Medical Inc.

Class II · Terminated

Class II FDA medical device recall by Bradshaw Medical Inc., initiated 2015-02-02, terminated 2020-01-10. FDA's recall record names no specific 510(k) clearance for this event. Product: Axial Torque Limiting Driver (1st generation) Product Usage: The Bradshaw Medical Axial Torque Limiting Driver is a reusable surgical instrument designed to limit the magnitude of applied (input) torque transferred to an external (output) device or fastener in the clockwise direction. The device is a non-adjustable mechanical click style hand torque tool which presents an audible and tactile click when the maximum desired torque is reached.. Reason: A component failure could cause a situation where the applied torque was significantly higher than the calibrated torque setting..

Recalling firm
Bradshaw Medical Inc.
Classification
Class II
Status
Terminated
Initiated
2015-02-02
Terminated
2020-01-10
Firm location
Kenosha, WI, United States

Product description

Axial Torque Limiting Driver (1st generation) Product Usage: The Bradshaw Medical Axial Torque Limiting Driver is a reusable surgical instrument designed to limit the magnitude of applied (input) torque transferred to an external (output) device or fastener in the clockwise direction. The device is a non-adjustable mechanical click style hand torque tool which presents an audible and tactile click when the maximum desired torque is reached.

Reason for recall

A component failure could cause a situation where the applied torque was significantly higher than the calibrated torque setting.