Recall Z-1632-2023— Siemens Healthcare Diagnostics, Inc.

Class II · Ongoing

Class II FDA medical device recall by Siemens Healthcare Diagnostics, Inc., initiated 2023-04-05. FDA's recall record names no specific 510(k) clearance for this event. Product: Atellica CH Toxicology Calibrator (TOX CAL)-The Atellica¿ CH Toxicology Calibrator (TOX CAL) is for in vitro diagnostic use in calibrating the Acet, ETOH and Sal assays using the Atellica¿ CH Analyzer Siemens Material Number (SMN): 11099440. Reason: Reassignment of the Atellica CH Toxicology Calibrators (TOX CAL) and ADVIA Chemistry ToxAmmonia Calibrators for the Atellica CH Salicylate (Sal) Assay due to a positive bias.

Recalling firm
Siemens Healthcare Diagnostics, Inc.
Classification
Class II
Status
Ongoing
Initiated
2023-04-05
Firm location
Tarrytown, NY, United States

Product description

Atellica CH Toxicology Calibrator (TOX CAL)-The Atellica¿ CH Toxicology Calibrator (TOX CAL) is for in vitro diagnostic use in calibrating the Acet, ETOH and Sal assays using the Atellica¿ CH Analyzer Siemens Material Number (SMN): 11099440

Reason for recall

Reassignment of the Atellica CH Toxicology Calibrators (TOX CAL) and ADVIA Chemistry ToxAmmonia Calibrators for the Atellica CH Salicylate (Sal) Assay due to a positive bias