Recall Z-1630-2021— Philips Respironics, Inc.

Class II · Terminated

Class II FDA medical device recall by Philips Respironics, Inc., initiated 2021-04-06, terminated 2024-01-17. FDA's recall record names no specific 510(k) clearance for this event. Product: BiPAP A40, Brazil: Continuous ventilator, non-life-supporting. Material Number: BR1111169. Reason: Certain devices were built with motor assemblies that could contain non-conforming plastic material. This could result in 1. Exposure to off-gassing of Volatile Organic Compounds not normally present creating a potential biosafety or toxicological hazard and 2. Sudden failure of the device causing a Ventilator Inoperative condition with the potential for asphyxia if not immediately identified and addressed by the care provider..

Recalling firm
Philips Respironics, Inc.
Classification
Class II
Status
Terminated
Initiated
2021-04-06
Terminated
2024-01-17
Firm location
Murrysville, PA, United States

Product description

BiPAP A40, Brazil: Continuous ventilator, non-life-supporting. Material Number: BR1111169

Reason for recall

Certain devices were built with motor assemblies that could contain non-conforming plastic material. This could result in 1. Exposure to off-gassing of Volatile Organic Compounds not normally present creating a potential biosafety or toxicological hazard and 2. Sudden failure of the device causing a Ventilator Inoperative condition with the potential for asphyxia if not immediately identified and addressed by the care provider.