Recall Z-1628-2014— Philips Medical Systems, Inc.
Class II · Terminated
Class II FDA medical device recall by Philips Medical Systems, Inc., initiated 2014-04-04, terminated 2017-11-07. It names one FDA 510(k) clearance: K110622. Product: Lithium Ion Battery M4605A and M4607A for use with IntelliVue Patient Monitors. The monitor is used for monitoring and recording of, and to generate alarms for, multiple physiological parameters.. Reason: The risk of battery failure increases with age, when a battery remains in use longer than 3 years after date of manufacture or 500 charge-discharge cycles. Such failure can result in overheating that in rare cases can cause the battery to ignite or explode..
- Recalling firm
- Philips Medical Systems, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2014-04-04
- Terminated
- 2017-11-07
- Firm location
- Andover, MA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Lithium Ion Battery M4605A and M4607A for use with IntelliVue Patient Monitors. The monitor is used for monitoring and recording of, and to generate alarms for, multiple physiological parameters.
Reason for recall
The risk of battery failure increases with age, when a battery remains in use longer than 3 years after date of manufacture or 500 charge-discharge cycles. Such failure can result in overheating that in rare cases can cause the battery to ignite or explode.