Recall Z-1628-2013— Surgical Instrument Service And Savings, Inc.
Class II · Terminated
Class II FDA medical device recall by Surgical Instrument Service And Savings, Inc., initiated 2013-05-03, terminated 2013-07-11. It names one FDA 510(k) clearance: K012669. Product: -90 Degree L-Hook Tip Arthroscopic Electrode, for use with 7205440 Switchpen, (Blue) 170mm. Ref. 7205441 Lot 122502 -Acromioplasty Electrode (Grey) 6.5" x 165mm. Ref. C5012A Lot 122288 -CAPSure 30 ArthroWand, CW, 30 Degree Shaft Angle, (Grey) 3.0mm. Ref. AC 1830-01 Lot 122494 -Coblator II Evac 70 Xtra Plasma Wand, w/Integrated Saline Line & Integrated Cable. Ref. EIC 5872-01 Lot 122295, 122308, 122626, 122325, -Coblator II, ENT ReFlex Ultra 45 w/Integrated Cable, (Blue). Ref. EIC 4845-01 Lot 122651, 122419, 122671, 122633 Electrosurgical cutting and coagulation device and accessories. -CoVac 50, Suction ArthroWand, w/Integrated Cable, (White/Grey). Ref. ASC 2530-01 Lot 122483, 122745 -CoVac 70, ICW ArthroWand 70 angle 3.0mm. Ref. ASC 3730-01 Lot 122483, -Eliminator, Right Angle ArthroWand (Beige). Ref. A 1345-01 Lot 122446 -Eliminator, Right Angle ArthroWand, 90 Degree w/Integrated Cable, (Grey) 4.5mm. Ref. AC 1345-01 Lot 122494, 122446 -ENT Coblator II PROcise XP Wand, w/Integrated Cable, Suction & Saline, (Blue). Ref. EIC 8872-01 Lot 122651, 122762 -Ligament Chisel Probe,Curved w/Integrated Cable, (Grey/Green). Ref. 7209647 Lot 122512 -LightWave Suction Ablator, 90 Degree, Hand Control, Multi-Angle, (Blue/White) 4.5mm x 142mm. Ref. IA-2000-S Lot 122512 -Lo-Pro 90 Degree ArthroWand, w/Integrated Cable, (Grey) 3.6mm. Ref. AC1336-01 Lot 122654, 122325 -Meniscectomy Electrode, Standard Design (Grey) 165mm. Ref. C5010A Lot 122765 -MeniVac ArthroWand w/Suction, w/Integrated Cable, (Grey) 3.0mm. Ref. ASC 5500-01 Lot 122630, 122494 -Microblastor 30 Wand, 30 Degree, w/Integrated Cable, (Grey) 0.9mm Tip. Ref. AC 40500-01 Lot 122293 -Right Angle ArthroWand, 90 Degree, w/Integrated Cable, (Grey) 3.5mm. Ref. AC 1340-01 Lot 123044, 122673 -Saber 30 ArthroWand, 30 Degree, w/Integrated Cable, (Grey) 3.0mm. Ref. AC 4330-01 Lot 122630, 122494, 123044, 122283, 122765 -SAPHYRE II Bipolar Ablation Probe, w/suction, High Profile 50 Degree, (Pewter). Ref. 7210112 Lot 122734 -SERFAS ENERGY Hook Probe 3.5mm. Ref. 279-350-501 Lot 123266 -Small Joint ArthroWand, Short Bevel 35, w/Integrated Cable, Suction (Grey/White) 2.3mm. Ref. AC 2823-01 Lot 122494, 122663 -StarVac ArthroWand, Short Bevel 35, w/Integrated Cable, Suction (Grey)5.5mm. Ref. ASC 4251-01 Lot 122630, 122593, 122483, 123044, 122283, 122595 -Super MultiVac 50, 50 Degree, w/Suction and Integrated Cable, (Grey/Black) 3.0mm. Ref. ASC 4830-01 Lot 122029, 123044, 122663 -Super TurboVac 90, 90 Degree Suction ArthroWand, (White/Grey) 3.5mm. Ref. ASC 4250-01 Lot 122029,122654, 122651, 122427, 122630, 122494, 122626, 122745, 122293, 122663, 122325, 122765 -TAC-C II Probe, w/Integrated Cable, (Blue/Grey). Ref. 7209628 Lot 122486 -TriStar 50 Suction ArthroWand, 50 Degree w/Integrated Cable, (White/Grey). Ref. ASC 4630-01 Lot 122283 -Turbinate, Tissue Coagulating Electrode, (White). Ref. 1120-4110-01 Lot 122486 -VAPR Hook Electrode W/Integrated Handpiece 3.5mm. Ref. 227305 Lot 122288, 122466 -VAPR LD, Suction Electrode, Large Diameter (Blue) 4.0mm. Ref. 225360 Lot 122512 -VAPR Premiere50-, w/Integrated Cable, Suction, (Black/White) 3.0mm. Ref. 227504 Lot 122029, 122455 -VAPR Premiere90 w/Suction tion, Integrated Cable, (Black/White) 3.3mm. Ref. 227204 Lot 122029, 122446, 122466, 122306, 122435, 122293, -VAPR S90 Suction Electrode, w/Integrated Cable, (Black/White). Ref. 225370 Lot 122577, 122455, 122593, 122288, 122466, 122589, 122356, 122639, 122325, 122318 -VAPR Side Effect Electrode (Blue) 3.5mm. Ref. 225301 Lot 122512 -VAPR Side Effect Electrode, w/Integrated Handpiece 3.5mm. Ref. 227301 Lot 122288, 122356, -VAPR Wedge Electrode (Short,) w/Integrated Cable (Black/White) 2.3mm x 85 mm. Ref. 227213 Lot 122409 -VAPR Wedge Electrode, w/Integrated Cable (Black/White) 2.3mm. Ref. 227203 Lot 122486, 122639 -VAPR Angled, Side Effect Electrode (Blue) 3.5mm. Ref. 225302 Lot 122512 -Vulcan Sa. Reason: The seal which maintains a sterile barrier for reprocessed medical devices became compromised to the point where product may fall out of the pouch..
- Recalling firm
- Surgical Instrument Service And Savings, Inc.
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2013-05-03
- Terminated
- 2013-07-11
- Firm location
- Redmond, OR, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
-90 Degree L-Hook Tip Arthroscopic Electrode, for use with 7205440 Switchpen, (Blue) 170mm. Ref. 7205441 Lot 122502 -Acromioplasty Electrode (Grey) 6.5" x 165mm. Ref. C5012A Lot 122288 -CAPSure 30 ArthroWand, CW, 30 Degree Shaft Angle, (Grey) 3.0mm. Ref. AC 1830-01 Lot 122494 -Coblator II Evac 70 Xtra Plasma Wand, w/Integrated Saline Line & Integrated Cable. Ref. EIC 5872-01 Lot 122295, 122308, 122626, 122325, -Coblator II, ENT ReFlex Ultra 45 w/Integrated Cable, (Blue). Ref. EIC 4845-01 Lot 122651, 122419, 122671, 122633 Electrosurgical cutting and coagulation device and accessories. -CoVac 50, Suction ArthroWand, w/Integrated Cable, (White/Grey). Ref. ASC 2530-01 Lot 122483, 122745 -CoVac 70, ICW ArthroWand 70 angle 3.0mm. Ref. ASC 3730-01 Lot 122483, -Eliminator, Right Angle ArthroWand (Beige). Ref. A 1345-01 Lot 122446 -Eliminator, Right Angle ArthroWand, 90 Degree w/Integrated Cable, (Grey) 4.5mm. Ref. AC 1345-01 Lot 122494, 122446 -ENT Coblator II PROcise XP Wand, w/Integrated Cable, Suction & Saline, (Blue). Ref. EIC 8872-01 Lot 122651, 122762 -Ligament Chisel Probe,Curved w/Integrated Cable, (Grey/Green). Ref. 7209647 Lot 122512 -LightWave Suction Ablator, 90 Degree, Hand Control, Multi-Angle, (Blue/White) 4.5mm x 142mm. Ref. IA-2000-S Lot 122512 -Lo-Pro 90 Degree ArthroWand, w/Integrated Cable, (Grey) 3.6mm. Ref. AC1336-01 Lot 122654, 122325 -Meniscectomy Electrode, Standard Design (Grey) 165mm. Ref. C5010A Lot 122765 -MeniVac ArthroWand w/Suction, w/Integrated Cable, (Grey) 3.0mm. Ref. ASC 5500-01 Lot 122630, 122494 -Microblastor 30 Wand, 30 Degree, w/Integrated Cable, (Grey) 0.9mm Tip. Ref. AC 40500-01 Lot 122293 -Right Angle ArthroWand, 90 Degree, w/Integrated Cable, (Grey) 3.5mm. Ref. AC 1340-01 Lot 123044, 122673 -Saber 30 ArthroWand, 30 Degree, w/Integrated Cable, (Grey) 3.0mm. Ref. AC 4330-01 Lot 122630, 122494, 123044, 122283, 122765 -SAPHYRE II Bipolar Ablation Probe, w/suction, High Profile 50 Degree, (Pewter). Ref. 7210112 Lot 122734 -SERFAS ENERGY Hook Probe 3.5mm. Ref. 279-350-501 Lot 123266 -Small Joint ArthroWand, Short Bevel 35, w/Integrated Cable, Suction (Grey/White) 2.3mm. Ref. AC 2823-01 Lot 122494, 122663 -StarVac ArthroWand, Short Bevel 35, w/Integrated Cable, Suction (Grey)5.5mm. Ref. ASC 4251-01 Lot 122630, 122593, 122483, 123044, 122283, 122595 -Super MultiVac 50, 50 Degree, w/Suction and Integrated Cable, (Grey/Black) 3.0mm. Ref. ASC 4830-01 Lot 122029, 123044, 122663 -Super TurboVac 90, 90 Degree Suction ArthroWand, (White/Grey) 3.5mm. Ref. ASC 4250-01 Lot 122029,122654, 122651, 122427, 122630, 122494, 122626, 122745, 122293, 122663, 122325, 122765 -TAC-C II Probe, w/Integrated Cable, (Blue/Grey). Ref. 7209628 Lot 122486 -TriStar 50 Suction ArthroWand, 50 Degree w/Integrated Cable, (White/Grey). Ref. ASC 4630-01 Lot 122283 -Turbinate, Tissue Coagulating Electrode, (White). Ref. 1120-4110-01 Lot 122486 -VAPR Hook Electrode W/Integrated Handpiece 3.5mm. Ref. 227305 Lot 122288, 122466 -VAPR LD, Suction Electrode, Large Diameter (Blue) 4.0mm. Ref. 225360 Lot 122512 -VAPR Premiere50-, w/Integrated Cable, Suction, (Black/White) 3.0mm. Ref. 227504 Lot 122029, 122455 -VAPR Premiere90 w/Suction tion, Integrated Cable, (Black/White) 3.3mm. Ref. 227204 Lot 122029, 122446, 122466, 122306, 122435, 122293, -VAPR S90 Suction Electrode, w/Integrated Cable, (Black/White). Ref. 225370 Lot 122577, 122455, 122593, 122288, 122466, 122589, 122356, 122639, 122325, 122318 -VAPR Side Effect Electrode (Blue) 3.5mm. Ref. 225301 Lot 122512 -VAPR Side Effect Electrode, w/Integrated Handpiece 3.5mm. Ref. 227301 Lot 122288, 122356, -VAPR Wedge Electrode (Short,) w/Integrated Cable (Black/White) 2.3mm x 85 mm. Ref. 227213 Lot 122409 -VAPR Wedge Electrode, w/Integrated Cable (Black/White) 2.3mm. Ref. 227203 Lot 122486, 122639 -VAPR Angled, Side Effect Electrode (Blue) 3.5mm. Ref. 225302 Lot 122512 -Vulcan Sa
Reason for recall
The seal which maintains a sterile barrier for reprocessed medical devices became compromised to the point where product may fall out of the pouch.