Recall Z-1623-2015— GE Healthcare
Class II · Terminated
Class II FDA medical device recall by GE Healthcare, initiated 2015-04-22, terminated 2015-07-15. It names one FDA 510(k) clearance: K013561. Product: CT LightSpeed 16 and Discovery ST Pet/CT DST. Catalogue # 2339985. Designed to be a head and whole body CT scanner utilizing a new solid state detector.. Reason: GE Healthcare has become aware that 3 bolts that secure a power supply to the rotating side of the gantry were over-torqued on certain Lightspeed 16 CT scanner or PET/CT DST Scanner.. In the unlikely event that the component becomes loose on the CT gantry during operation, it could result in serious bodily injury if the component were expelled. There were no incidents or injuries reported as a res.
- Recalling firm
- GE Healthcare
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2015-04-22
- Terminated
- 2015-07-15
- Firm location
- Waukesha, WI, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
CT LightSpeed 16 and Discovery ST Pet/CT DST. Catalogue # 2339985. Designed to be a head and whole body CT scanner utilizing a new solid state detector.
Reason for recall
GE Healthcare has become aware that 3 bolts that secure a power supply to the rotating side of the gantry were over-torqued on certain Lightspeed 16 CT scanner or PET/CT DST Scanner.. In the unlikely event that the component becomes loose on the CT gantry during operation, it could result in serious bodily injury if the component were expelled. There were no incidents or injuries reported as a res