Recall Z-1622-2020— Respironics California, LLC
Class II · Ongoing
Class II FDA medical device recall by Respironics California, LLC, initiated 2020-03-05. It names one FDA 510(k) clearance: K082660. Product: Philips V60 Ventilators with Power Management PCBA part number 1055906. Reason: Solder connection failure on Power Management printed circuit board assembly of ventilators may cause 1) Blower to lose power, spool down, and trigger visual and audible High Priority Check Vent alarm indicating a need for alternative ventilation, or 2) Intermittent connection may disrupt expected operation and trigger unexpected shutdown without alarm. Failures may lead to hypercarbia/hypoxemia.
- Recalling firm
- Respironics California, LLC
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2020-03-05
- Firm location
- Carlsbad, CA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Philips V60 Ventilators with Power Management PCBA part number 1055906
Reason for recall
Solder connection failure on Power Management printed circuit board assembly of ventilators may cause 1) Blower to lose power, spool down, and trigger visual and audible High Priority Check Vent alarm indicating a need for alternative ventilation, or 2) Intermittent connection may disrupt expected operation and trigger unexpected shutdown without alarm. Failures may lead to hypercarbia/hypoxemia
Data sourced from openFDA. This site is unofficial and independent of the FDA.