Recall Z-1621-2021— Boston Scientific Corporation

Class II · Terminated

Class II FDA medical device recall by Boston Scientific Corporation, initiated 2021-04-21, terminated 2023-08-07. FDA's recall record names no specific 510(k) clearance for this event. Product: ELUVIA Drug-Eluting Vascular Stent System - Product Usage: indicated for improving luminal diameter in the treatment of symptomatic de-novo or restenotic lesions in the native superficial femoral artery (SFA) and/or proximal popliteal artery with reference vessel diameters (RVD) ranging from 4.0-6.0 mm and total lesion lengths up to 190 mm.. Reason: One batch of the ELUVIATM Drug Eluting Vascular Stent System incorrect placement of some 6x40mm ELUVIA stents into 6x100mm packaging..

Recalling firm
Boston Scientific Corporation
Classification
Class II
Status
Terminated
Initiated
2021-04-21
Terminated
2023-08-07
Firm location
Maple Grove, MN, United States

Product description

ELUVIA Drug-Eluting Vascular Stent System - Product Usage: indicated for improving luminal diameter in the treatment of symptomatic de-novo or restenotic lesions in the native superficial femoral artery (SFA) and/or proximal popliteal artery with reference vessel diameters (RVD) ranging from 4.0-6.0 mm and total lesion lengths up to 190 mm.

Reason for recall

One batch of the ELUVIATM Drug Eluting Vascular Stent System incorrect placement of some 6x40mm ELUVIA stents into 6x100mm packaging.

Data sourced from openFDA. This site is unofficial and independent of the FDA.