Recall Z-1621-2015— bioMerieux, Inc.

Class II · Terminated

Class II FDA medical device recall by bioMerieux, Inc., initiated 2015-04-14, terminated 2017-12-13. FDA's recall record names no specific 510(k) clearance for this event. Product: MYLA server HP Proliant: DL380-G8 and ML350-G6 version 3.xx computer application software. MYLA V3 is a computer application ("Middleware") based on Web 2.0 technology which: Interfaces between the instruments connected to the application and the LIS(s) (Laboratory Information System(s).. Reason: The MYLA¿ server could slow down due to the volume (weight) of the data to manage and it could also stop processing, impacting the server start-up phase. The problem occurs in high volume settings. The MYLA server slows down and will stop processing completely if any task requires more than 5 minutes to complete. The server then fails to reboot and requires service, provided remotely or on-site.

Recalling firm
bioMerieux, Inc.
Classification
Class II
Status
Terminated
Initiated
2015-04-14
Terminated
2017-12-13
Firm location
Durham, NC, United States

Product description

MYLA server HP Proliant: DL380-G8 and ML350-G6 version 3.xx computer application software. MYLA V3 is a computer application ("Middleware") based on Web 2.0 technology which: Interfaces between the instruments connected to the application and the LIS(s) (Laboratory Information System(s).

Reason for recall

The MYLA¿ server could slow down due to the volume (weight) of the data to manage and it could also stop processing, impacting the server start-up phase. The problem occurs in high volume settings. The MYLA server slows down and will stop processing completely if any task requires more than 5 minutes to complete. The server then fails to reboot and requires service, provided remotely or on-site

Data sourced from openFDA. This site is unofficial and independent of the FDA.