Recall Z-1618-2020— Karl Storz Endoscopy

Class II · Terminated

Class II FDA medical device recall by Karl Storz Endoscopy, initiated 2019-12-06, terminated 2022-07-01. FDA's recall record names no specific 510(k) clearance for this event. Product: Karl Storz SE & Co. KG, C-MAC Video Laryngoscope MIL # 1, for the following REF numbers: 1. 8403xx series, which includes: 8403AX, AXC, BX, BXC, DXC, EXC, GXC, HX, HXK, HXP, KXC, MXC 2. 8401xx series, which includes: 8401AX, AXC, BX, BXC, DXC, GXC, HX, HXP, KXC. Product Usage: Video Laryngoscopes are used to visualize the airways and vocal cords during endotracheal intubation and for the inspection and examination of the upper respiratory tract. Video Laryngoscopes are designed for transient use in invasive procedures through a body orifice.. Reason: The firm has become aware that there is a potential that the glass cover at the distal end of the camera/led lighting on the video laryngoscope could become loose during reprocessing or use. If this occurs during usage, it may result in a delay in procedure and/or require removal from body..

Recalling firm
Karl Storz Endoscopy
Classification
Class II
Status
Terminated
Initiated
2019-12-06
Terminated
2022-07-01
Firm location
El Segundo, CA, United States

Product description

Karl Storz SE & Co. KG, C-MAC Video Laryngoscope MIL # 1, for the following REF numbers: 1. 8403xx series, which includes: 8403AX, AXC, BX, BXC, DXC, EXC, GXC, HX, HXK, HXP, KXC, MXC 2. 8401xx series, which includes: 8401AX, AXC, BX, BXC, DXC, GXC, HX, HXP, KXC. Product Usage: Video Laryngoscopes are used to visualize the airways and vocal cords during endotracheal intubation and for the inspection and examination of the upper respiratory tract. Video Laryngoscopes are designed for transient use in invasive procedures through a body orifice.

Reason for recall

The firm has become aware that there is a potential that the glass cover at the distal end of the camera/led lighting on the video laryngoscope could become loose during reprocessing or use. If this occurs during usage, it may result in a delay in procedure and/or require removal from body.

Data sourced from openFDA. This site is unofficial and independent of the FDA.