Recall Z-1615-2021— Medtronic Inc.

Class II · Ongoing

Class II FDA medical device recall by Medtronic Inc., initiated 2021-02-26. FDA's recall record names no specific 510(k) clearance for this event. Product: Medtronic REF: MMT-7020D4 Guardian Sensor (3) for the following packaging: GS3 1PK 2L CANADA/ GTIN: 00763000179595¿ Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. *Product not distributed within US.. Reason: Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings..

Recalling firm
Medtronic Inc.
Classification
Class II
Status
Ongoing
Initiated
2021-02-26
Firm location
Northridge, CA, United States

Product description

Medtronic REF: MMT-7020D4 Guardian Sensor (3) for the following packaging: GS3 1PK 2L CANADA/ GTIN: 00763000179595¿ Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. *Product not distributed within US.

Reason for recall

Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings.

Data sourced from openFDA. This site is unofficial and independent of the FDA.