Recall Z-1611-2021— Medtronic Inc.

Class II · Ongoing

Class II FDA medical device recall by Medtronic Inc., initiated 2021-02-26. FDA's recall record names no specific 510(k) clearance for this event. Product: Medtronic REF: MMT-7020C5 Guardian Sensor (3) for the following packaging: GS3 5PK 1L AUS/ GTIN: 20763000252056 GS3 5PK 1L AUS/ GTIN: 20763000414362 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. *Product not distributed within US.. Reason: Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings..

Recalling firm
Medtronic Inc.
Classification
Class II
Status
Ongoing
Initiated
2021-02-26
Firm location
Northridge, CA, United States

Product description

Medtronic REF: MMT-7020C5 Guardian Sensor (3) for the following packaging: GS3 5PK 1L AUS/ GTIN: 20763000252056 GS3 5PK 1L AUS/ GTIN: 20763000414362 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. *Product not distributed within US.

Reason for recall

Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings.

Data sourced from openFDA. This site is unofficial and independent of the FDA.