Recall Z-1608-2026— I.T.S. GmbH

Class II · Ongoing

Class II FDA medical device recall by I.T.S. GmbH, initiated 2026-01-23. It names one FDA 510(k) clearance: K210935. Product: Brand Name: I.T.S. Pelvic Reconstruction System (PRS RX & PRS Phoenix) with the below descriptions and corresponding article numbers. 1. Anterior Column Plate; Article Numbers: 21218-16, 21219-16. 2. Anterior Column Plate; Article Numbers: 21218-14, 21219-14. 3. Biplanar 2-Column Plate; Article Numbers: 21216-10, 21216-12, 21217-10, 21217-12. 4. Cancellous Screw; Article Numbers: 30591-36, 30591-70, 30591-75, 30591-80, 30591-85, 30591-90. 5. Cancellous Screw, Locking; Article Numbers: 37422-12-N, 37422-14-N, 37422-16-N, 37422-18-N, 37422-20-N, 37422-22-N, 37422-24-N, 37422-26-N, 37422-28-N, 37422-30-N, 37422-32-N, 37422-34-N, 37422-36-N, 37422-38-N, 37422-40-N, 37422-42-N, 37422-44-N, 37422-46-N, 37422-48-N, 37422-50-N, 37422-55-N, 37422-60-N, 37422-65-N, 37422-70-N, 37422-75-N, 37422-80-N, 37422-85-N, 37422-90-N, 37422-95-N, 37422-100-N, 37422-105-N, 37422-110-N, 37422-115-N, 37422-120-N, 37592-18, 37592-20, 37592-22, 37592-24, 37592-26, 37592-28, 37592-30, 37592-32, 37592-34, 37592-36, 37592-38, 37592-40, 37592-42, 37592-46, 37592-48, 37592-50, 37592-54, 37592-56, 37592-58, 37592-60, 37592-65, 37592-70, 37592-75, 37592-80, 37592-85, 37592-90. 6. Cannulated Screw, Cancellous; Article Numbers: 31851-60, 31851-65, 31851-70, 31851-75, 31851-80, 31851-85, 31851-90, 31851-95, 31851-100, 31851-105, 31851-110, 31851-115, 31851-120, 31851-125, 31851-130, 31851-135, 31851-140, 31851-145, 31851-150, 31851-155, 31851-160, 31851-165, 31851-170, 31851-175, 31851-180, 31851-185, 31851-190, 31851-195, 31851-200, 31854-60, 31854-65, 31854-70, 31854-75, 31854-80, 31854-85, 31854-90, 31854-95, 31854-100, 31854-105, 31854-110, 31854-115, 31854-120, 31854-125, 31854-130, 31854-135, 31854-140, 31854-145, 31854-150, 31854-155, 31854-160, 31854-165, 31854-170, 31854-175, 31854-180, 31854-185, 31854-190, 31854-195, 31854-200. 7. Cortical Screw; Article Numbers: 32351-10, 32351-12, 32351-14, 32351-16, 32351-18, 32351-20, 32351-22, 32351-24, 32351-26, 32351-28, 32351-30, 32351-32, 32351-34, 32351-36, 32351-38, 32351-40, 32351-42, 32351-44, 32351-46, 32351-48, 32351-50, 32351-55, 32351-60, 32351-65, 32351-70, 32351-75, 32351-80, 32351-85, 32352-16, 32352-18, 32352-20, 32352-22, 32352-24, 32352-26, 32352-28, 32352-30, 32352-32, 32352-34, 32352-36, 32352-38, 32352-40, 32352-42, 32352-44, 32352-46, 32352-48, 32352-50, 32352-55, 32352-60, 32352-65, 32352-70, 32352-75, 32352-80, 32352-85, 32352-90, 32352-95, 32352-100, 32352-105, 32352-110, 32352-115, 32352-120, 32455-18, 32455-20, 32455-22, 32455-24, 32455-26, 32455-28, 32455-30, 32455-32, 32455-34, 32455-36, 32455-38, 32455-40, 32455-42, 32455-44, 32455-46, 32455-48, 32455-50, 32455-52, 32455-54, 32455-56, 32455-58, 32455-60, 32455-65, 32455-70, 32455-75, 32455-80, 32455-85, 32455-90, 32475-32. 8. Cortical Screw, Locking; Article Numbers: 37455-18, 37455-20, 37455-22, 37455-24, 37455-26, 37455-28, 37455-30, 37455-32, 37455-34, 37455-36, 37455-38, 37455-40, 37455-42, 37455-44, 37455-46, 37455-48, 37455-50, 37455-52, 37455-54, 37455-56, 37455-60, 37455-65, 37455-70, 37455-75, 37455-80, 37455-85, 37455-90. 9. Curved Plate; Article Numbers: 211911-4, 211911-5, 211911-6, 211911-7, 211911-8, 211911-10, 211911-12, 211911-14, 211911-16, 211911-20. 10. Posterior Column Plate, Left; Article Numbers: 212121-6. 11. Posterior Column Plate, Right; Article Numbers: 212131-6. 12. Posterior Wall Plate, Extended, Left; Article Numbers: 212101-SM. 13. Posterior Wall Plate, Extended, Left; Article Numbers: 212101-LA. 14. Posterior Wall Plate, Extended, Right; Article Numbers: 212111-LA, 212111-SM. 15. Posterior Wall Plate, Left; Article Numbers: 212141-7, 212141-8. 16. Posterior Wall Plate, Right; Article Numbers: 212151-7, 212151-8. 17. PRS Plate; Article Numbers: 21194-6, 21194-7, 21194-8, 21195-10, 21195-12, 21195-7. 18. Quadrilateral Buttress Plate; Article Numbers: 21208-13, 21209-13. 19. Quadrilateral Buttress Plate; Article Numbers: 21208-11, 21209-11. 20. Rim Plate, Left; Article. Reason: Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU)..

Recalling firm
I.T.S. GmbH
Classification
Class II
Status
Ongoing
Initiated
2026-01-23
Firm location
Lasnitzhohe, Austria

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Brand Name: I.T.S. Pelvic Reconstruction System (PRS RX & PRS Phoenix) with the below descriptions and corresponding article numbers. 1. Anterior Column Plate; Article Numbers: 21218-16, 21219-16. 2. Anterior Column Plate; Article Numbers: 21218-14, 21219-14. 3. Biplanar 2-Column Plate; Article Numbers: 21216-10, 21216-12, 21217-10, 21217-12. 4. Cancellous Screw; Article Numbers: 30591-36, 30591-70, 30591-75, 30591-80, 30591-85, 30591-90. 5. Cancellous Screw, Locking; Article Numbers: 37422-12-N, 37422-14-N, 37422-16-N, 37422-18-N, 37422-20-N, 37422-22-N, 37422-24-N, 37422-26-N, 37422-28-N, 37422-30-N, 37422-32-N, 37422-34-N, 37422-36-N, 37422-38-N, 37422-40-N, 37422-42-N, 37422-44-N, 37422-46-N, 37422-48-N, 37422-50-N, 37422-55-N, 37422-60-N, 37422-65-N, 37422-70-N, 37422-75-N, 37422-80-N, 37422-85-N, 37422-90-N, 37422-95-N, 37422-100-N, 37422-105-N, 37422-110-N, 37422-115-N, 37422-120-N, 37592-18, 37592-20, 37592-22, 37592-24, 37592-26, 37592-28, 37592-30, 37592-32, 37592-34, 37592-36, 37592-38, 37592-40, 37592-42, 37592-46, 37592-48, 37592-50, 37592-54, 37592-56, 37592-58, 37592-60, 37592-65, 37592-70, 37592-75, 37592-80, 37592-85, 37592-90. 6. Cannulated Screw, Cancellous; Article Numbers: 31851-60, 31851-65, 31851-70, 31851-75, 31851-80, 31851-85, 31851-90, 31851-95, 31851-100, 31851-105, 31851-110, 31851-115, 31851-120, 31851-125, 31851-130, 31851-135, 31851-140, 31851-145, 31851-150, 31851-155, 31851-160, 31851-165, 31851-170, 31851-175, 31851-180, 31851-185, 31851-190, 31851-195, 31851-200, 31854-60, 31854-65, 31854-70, 31854-75, 31854-80, 31854-85, 31854-90, 31854-95, 31854-100, 31854-105, 31854-110, 31854-115, 31854-120, 31854-125, 31854-130, 31854-135, 31854-140, 31854-145, 31854-150, 31854-155, 31854-160, 31854-165, 31854-170, 31854-175, 31854-180, 31854-185, 31854-190, 31854-195, 31854-200. 7. Cortical Screw; Article Numbers: 32351-10, 32351-12, 32351-14, 32351-16, 32351-18, 32351-20, 32351-22, 32351-24, 32351-26, 32351-28, 32351-30, 32351-32, 32351-34, 32351-36, 32351-38, 32351-40, 32351-42, 32351-44, 32351-46, 32351-48, 32351-50, 32351-55, 32351-60, 32351-65, 32351-70, 32351-75, 32351-80, 32351-85, 32352-16, 32352-18, 32352-20, 32352-22, 32352-24, 32352-26, 32352-28, 32352-30, 32352-32, 32352-34, 32352-36, 32352-38, 32352-40, 32352-42, 32352-44, 32352-46, 32352-48, 32352-50, 32352-55, 32352-60, 32352-65, 32352-70, 32352-75, 32352-80, 32352-85, 32352-90, 32352-95, 32352-100, 32352-105, 32352-110, 32352-115, 32352-120, 32455-18, 32455-20, 32455-22, 32455-24, 32455-26, 32455-28, 32455-30, 32455-32, 32455-34, 32455-36, 32455-38, 32455-40, 32455-42, 32455-44, 32455-46, 32455-48, 32455-50, 32455-52, 32455-54, 32455-56, 32455-58, 32455-60, 32455-65, 32455-70, 32455-75, 32455-80, 32455-85, 32455-90, 32475-32. 8. Cortical Screw, Locking; Article Numbers: 37455-18, 37455-20, 37455-22, 37455-24, 37455-26, 37455-28, 37455-30, 37455-32, 37455-34, 37455-36, 37455-38, 37455-40, 37455-42, 37455-44, 37455-46, 37455-48, 37455-50, 37455-52, 37455-54, 37455-56, 37455-60, 37455-65, 37455-70, 37455-75, 37455-80, 37455-85, 37455-90. 9. Curved Plate; Article Numbers: 211911-4, 211911-5, 211911-6, 211911-7, 211911-8, 211911-10, 211911-12, 211911-14, 211911-16, 211911-20. 10. Posterior Column Plate, Left; Article Numbers: 212121-6. 11. Posterior Column Plate, Right; Article Numbers: 212131-6. 12. Posterior Wall Plate, Extended, Left; Article Numbers: 212101-SM. 13. Posterior Wall Plate, Extended, Left; Article Numbers: 212101-LA. 14. Posterior Wall Plate, Extended, Right; Article Numbers: 212111-LA, 212111-SM. 15. Posterior Wall Plate, Left; Article Numbers: 212141-7, 212141-8. 16. Posterior Wall Plate, Right; Article Numbers: 212151-7, 212151-8. 17. PRS Plate; Article Numbers: 21194-6, 21194-7, 21194-8, 21195-10, 21195-12, 21195-7. 18. Quadrilateral Buttress Plate; Article Numbers: 21208-13, 21209-13. 19. Quadrilateral Buttress Plate; Article Numbers: 21208-11, 21209-11. 20. Rim Plate, Left; Article

Reason for recall

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).