Recall Z-1601-2026— I.T.S. GmbH

Class II · Ongoing

Class II FDA medical device recall by I.T.S. GmbH, initiated 2026-01-23. It names one FDA 510(k) clearance: K063672. Product: I.T.S. Fibula Plate PROlock with Angular Stability with the below descriptions and corresponding article numbers. 1. Distal Tubular Plate; Article Numbers: 21602-3, 21602-4, 21602-5, 21602-6, 21602-7, 21602-8, 21603-3, 21603-4, 21603-5, 21603-6, 21603-7, 21603-8. 2. Fibula Plate 3.5mm; Article Numbers: 21225-4, 21225-6, 21225-8, 21225-10, 21225-12, 21226-4, 21226-6, 21226-8, 21226-10, 21226-12. 3. Universal Tubular Plate; Article Numbers: 21601-5, 21601-6, 21601-7, 21601-8, 21601-10, 21601-12, 21601-14, 21601-16.. Reason: Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU)..

Recalling firm
I.T.S. GmbH
Classification
Class II
Status
Ongoing
Initiated
2026-01-23
Firm location
Lasnitzhohe, Austria

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

I.T.S. Fibula Plate PROlock with Angular Stability with the below descriptions and corresponding article numbers. 1. Distal Tubular Plate; Article Numbers: 21602-3, 21602-4, 21602-5, 21602-6, 21602-7, 21602-8, 21603-3, 21603-4, 21603-5, 21603-6, 21603-7, 21603-8. 2. Fibula Plate 3.5mm; Article Numbers: 21225-4, 21225-6, 21225-8, 21225-10, 21225-12, 21226-4, 21226-6, 21226-8, 21226-10, 21226-12. 3. Universal Tubular Plate; Article Numbers: 21601-5, 21601-6, 21601-7, 21601-8, 21601-10, 21601-12, 21601-14, 21601-16.

Reason for recall

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Data sourced from openFDA. This site is unofficial and independent of the FDA.