Recall Z-1600-2024— Siemens Medical Solutions USA, Inc.

Class II · Ongoing

Class II FDA medical device recall by Siemens Medical Solutions USA, Inc., initiated 2024-02-23. It names one FDA 510(k) clearance: K210743. Product: ACUSON Redwood 1.0 Diagnostic Ultrasound System, REF: 11503314, running software version VA10A to VA10G. Reason: On ultrasound systems, when Cardiac DICOM SR feature is configured to display either minimum or maximum measured value, and multiple cardiac region measurements are made, and results are exported into the SR feature, then The SR viewer will display the LAST measured value, not min or max, which could contribute to patient condition misdiagnosis or negatively influence patient management decisions..

Recalling firm
Siemens Medical Solutions USA, Inc.
Classification
Class II
Status
Ongoing
Initiated
2024-02-23
Firm location
Issaquah, WA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

ACUSON Redwood 1.0 Diagnostic Ultrasound System, REF: 11503314, running software version VA10A to VA10G

Reason for recall

On ultrasound systems, when Cardiac DICOM SR feature is configured to display either minimum or maximum measured value, and multiple cardiac region measurements are made, and results are exported into the SR feature, then The SR viewer will display the LAST measured value, not min or max, which could contribute to patient condition misdiagnosis or negatively influence patient management decisions.

Data sourced from openFDA. This site is unofficial and independent of the FDA.