Recall Z-1598-2024— Siemens Medical Solutions USA, Inc
Class II · Ongoing
Class II FDA medical device recall by Siemens Medical Solutions USA, Inc, initiated 2024-03-21. It names one FDA 510(k) clearance: K181407. Product: Siemens Artis Models: Interventional Fluoroscopic X-Ray Systems-Models: Artis zee floor- 10094135 Artis zee ceiling- 10094137 Artis zee multi-purpose- 10094139 Artis zee biplane- 10094141 Artis zeego- 10280959 Artis Q floor- 10848280 Artis Q ceiling- 10848281 Artis Q biplane- 10848282 Artis Q zeego- 10848283 Artis Q.zen floor- 10848353 Artis Q.zen ceiling- 10848354 Artis Q.zen biplane- 10848355 Artis zee floor MN- 10094142 Artis zee biplane MN- 10094143. Reason: Coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message TUBE HOT, have a break . the system cannot be operated normally, may result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system.
- Recalling firm
- Siemens Medical Solutions USA, Inc
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2024-03-21
- Firm location
- Malvern, PA, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Siemens Artis Models: Interventional Fluoroscopic X-Ray Systems-Models: Artis zee floor- 10094135 Artis zee ceiling- 10094137 Artis zee multi-purpose- 10094139 Artis zee biplane- 10094141 Artis zeego- 10280959 Artis Q floor- 10848280 Artis Q ceiling- 10848281 Artis Q biplane- 10848282 Artis Q zeego- 10848283 Artis Q.zen floor- 10848353 Artis Q.zen ceiling- 10848354 Artis Q.zen biplane- 10848355 Artis zee floor MN- 10094142 Artis zee biplane MN- 10094143
Reason for recall
Coolant level in the cooling circuit drops below a certain level, this may result in a situation in which the X-ray tube is no longer sufficiently cooled and the system will display the message TUBE HOT, have a break . the system cannot be operated normally, may result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system
Data sourced from openFDA. This site is unofficial and independent of the FDA.