Recall Z-1598-2013— Toshiba American Medical Systems Inc

Class II · Terminated

Class II FDA medical device recall by Toshiba American Medical Systems Inc, initiated 2013-04-03, terminated 2016-02-05. It names one FDA 510(k) clearance: K110785. Product: TOSHIBA Kalare Diagnostic X-Ray System K110785. Reason: Firm initiated recall because when radiographic conditions with the small focus are set and the radiography ready switch is held down, the large focus is also heated after a few seconds or more. This occurs when the Automatic Exposure Function (AEC) function is used. As a result, a radiographic tube current exceeding the set value flows and a brighter image than expected is acquired..

Recalling firm
Toshiba American Medical Systems Inc
Classification
Class II
Status
Terminated
Initiated
2013-04-03
Terminated
2016-02-05
Firm location
Tustin, CA, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

TOSHIBA Kalare Diagnostic X-Ray System K110785

Reason for recall

Firm initiated recall because when radiographic conditions with the small focus are set and the radiography ready switch is held down, the large focus is also heated after a few seconds or more. This occurs when the Automatic Exposure Function (AEC) function is used. As a result, a radiographic tube current exceeding the set value flows and a brighter image than expected is acquired.