Recall Z-1597-2019— Maquet Cardiovascular, LLC
Class II · Terminated
Class II FDA medical device recall by Maquet Cardiovascular, LLC, initiated 2019-04-18, terminated 2025-07-28. It names one FDA 510(k) clearance: K080592. Product: Maquet Getinge-BEQ-T 30914 ECC Pump Pack 1/4" Material: 709000011. Reason: Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured.
- Recalling firm
- Maquet Cardiovascular, LLC
- Classification
- Class II
- Status
- Terminated
- Initiated
- 2019-04-18
- Terminated
- 2025-07-28
- Firm location
- Wayne, NJ, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Maquet Getinge-BEQ-T 30914 ECC Pump Pack 1/4" Material: 709000011
Reason for recall
Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured
Data sourced from openFDA. This site is unofficial and independent of the FDA.