Recall Z-1596-2021— Roche Diagnostics Operations, Inc.
Class II · Ongoing
Class II FDA medical device recall by Roche Diagnostics Operations, Inc., initiated 2021-04-07. It names one FDA 510(k) clearance: K082699. Product: Elecsys Troponin I Immunoassay: Catalog# 05094810160 - Product Usage: intended to aid in the diagnosis of myocardial infarction.. Reason: Recall reason is possible sporadic false negative conventional troponin results. Roche has confirmed an issue with running Elecsys Troponin I/Troponin I STAT and Elecsys proBNP II/proBNP II STAT on the cobas e 601 and 602 modules. When the proBNP II analyte is measured immediately before the Troponin I analyte on the same module, a decreased recovery on Troponin I was observed..
- Recalling firm
- Roche Diagnostics Operations, Inc.
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2021-04-07
- Firm location
- Indianapolis, IN, United States
510(k) clearances named in this recall
1 on FDA's recordK-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.
Product description
Elecsys Troponin I Immunoassay: Catalog# 05094810160 - Product Usage: intended to aid in the diagnosis of myocardial infarction.
Reason for recall
Recall reason is possible sporadic false negative conventional troponin results. Roche has confirmed an issue with running Elecsys Troponin I/Troponin I STAT and Elecsys proBNP II/proBNP II STAT on the cobas e 601 and 602 modules. When the proBNP II analyte is measured immediately before the Troponin I analyte on the same module, a decreased recovery on Troponin I was observed.