Recall Z-1595-2023— Steris Corporation
Class II · Ongoing
Class II FDA medical device recall by Steris Corporation, initiated 2023-04-10. FDA's recall record names no specific 510(k) clearance for this event. Product: Reliance Vision Multi-Chamber Washer/Disinfector, Reference Number FH83062. Reason: In the remote occurrence in which the electrical contactor component present in the drying chambers of the Reliance Vision Multi-Chamber Washer/Disinfector malfunctions, the heating elements in the drying chamber could overheat, eventually resulting in smoke and/or fire..
- Recalling firm
- Steris Corporation
- Classification
- Class II
- Status
- Ongoing
- Initiated
- 2023-04-10
- Firm location
- Mentor, OH, United States
Product description
Reliance Vision Multi-Chamber Washer/Disinfector, Reference Number FH83062
Reason for recall
In the remote occurrence in which the electrical contactor component present in the drying chambers of the Reliance Vision Multi-Chamber Washer/Disinfector malfunctions, the heating elements in the drying chamber could overheat, eventually resulting in smoke and/or fire.