Recall Z-1594-2024— Randox Laboratories Ltd.

Class II · Ongoing

Class II FDA medical device recall by Randox Laboratories Ltd., initiated 2024-01-31. It names one FDA 510(k) clearance: K002674. Product: Microalbumin Calibrator Series (mALB CAL)-IVD to calibrate the Microalbumin Assay Catalog Number: MA2426. Reason: Microalbumin Calibrator Series, MA1567 & MA2426, is running with a negative bias compared to other method may misclassify patient results. Users may experience a shift in Quality Control and patient sample recovery of up to +12%..

Recalling firm
Randox Laboratories Ltd.
Classification
Class II
Status
Ongoing
Initiated
2024-01-31
Firm location
Crumlin (North), Ireland

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

Microalbumin Calibrator Series (mALB CAL)-IVD to calibrate the Microalbumin Assay Catalog Number: MA2426

Reason for recall

Microalbumin Calibrator Series, MA1567 & MA2426, is running with a negative bias compared to other method may misclassify patient results. Users may experience a shift in Quality Control and patient sample recovery of up to +12%.