Recall Z-1582-2013— Roche Diagnostics Operations, Inc.

Class II · Terminated

Class II FDA medical device recall by Roche Diagnostics Operations, Inc., initiated 2013-05-16, terminated 2014-04-16. It names one FDA 510(k) clearance: K093460. Product: CoaguChek XS Pro System The CoaguChek XS Pro System (CoaguChek XS Pro meter and CoaguChek XS PT Test strips) quantitatively determines prothrombin time ("PT"), using capillary blood or whole blood from a vein (non anticoagulated venous whole blood).. Reason: Roche has confirmed the potential for an undetected elevated INR test result with the CoaguChek XS, CoaguChek XS Plus, and CoaguChek XS Pro meters. In rare cases, instead of a value, an ERROR 6 message is displayed..

Recalling firm
Roche Diagnostics Operations, Inc.
Classification
Class II
Status
Terminated
Initiated
2013-05-16
Terminated
2014-04-16
Firm location
Indianapolis, IN, United States

510(k) clearances named in this recall

1 on FDA's record

K-numbers as listed on FDA's Recall Enterprise System record for this event (openFDA device/recall k_numbers). The recall may also cover devices cleared under other submissions not named here.

Product description

CoaguChek XS Pro System The CoaguChek XS Pro System (CoaguChek XS Pro meter and CoaguChek XS PT Test strips) quantitatively determines prothrombin time ("PT"), using capillary blood or whole blood from a vein (non anticoagulated venous whole blood).

Reason for recall

Roche has confirmed the potential for an undetected elevated INR test result with the CoaguChek XS, CoaguChek XS Plus, and CoaguChek XS Pro meters. In rare cases, instead of a value, an ERROR 6 message is displayed.

Data sourced from openFDA. This site is unofficial and independent of the FDA.