Recall Z-1580-2013— Medtronic Neuromodulation

Class I · Terminated

Class I FDA medical device recall by Medtronic Neuromodulation, initiated 2013-06-03, terminated 2015-11-16. FDA's recall record names no specific 510(k) clearance for this event. Product: Medtronic SynchroMed EL Programmable pumps, Models 8626, 8626L, 8627, 8627L, (The SynchroMed EL Pumps are supplied in 10 ml or 18 ml reservoir size.) Sterilized using ethylene oxide. The pump is part of an infusion system that stores and delivers a prescribed drug to a specific site. The drug is stored in the pump reservoir. Per a programmed prescription, the drug moves from the pump reservoir, through the pump tubing, catheter port, and catheter, to the infusion site. The SynchroMed II Programmable Pump is indicated when patient therapy requires the chronic infusion of the following drugs or fluids:. Reason: Medtronic Neuromodulation is providing Healthcare Providers with safety information and patient management recommendations related to the potential for electrical shorting internal to they SynchroMed infusion pump. An electrical short circuit in a feedthrough may present as a motor stall or low battery reset/alarm and lead to a loss of or reduction in therapy which may result in the return of und.

Recalling firm
Medtronic Neuromodulation
Classification
Class I
Status
Terminated
Initiated
2013-06-03
Terminated
2015-11-16
Firm location
Minneapolis, MN, United States

Product description

Medtronic SynchroMed EL Programmable pumps, Models 8626, 8626L, 8627, 8627L, (The SynchroMed EL Pumps are supplied in 10 ml or 18 ml reservoir size.) Sterilized using ethylene oxide. The pump is part of an infusion system that stores and delivers a prescribed drug to a specific site. The drug is stored in the pump reservoir. Per a programmed prescription, the drug moves from the pump reservoir, through the pump tubing, catheter port, and catheter, to the infusion site. The SynchroMed II Programmable Pump is indicated when patient therapy requires the chronic infusion of the following drugs or fluids:

Reason for recall

Medtronic Neuromodulation is providing Healthcare Providers with safety information and patient management recommendations related to the potential for electrical shorting internal to they SynchroMed infusion pump. An electrical short circuit in a feedthrough may present as a motor stall or low battery reset/alarm and lead to a loss of or reduction in therapy which may result in the return of und

Data sourced from openFDA. This site is unofficial and independent of the FDA.